Specifications
The 8350 pH and 8351 Redox probes are dedicated for process measurements. The pH probes
incorporate impedance control and a temperature sensor in contact with the sample.
8350.4 8350.5 8351
1 - Glass
2 - Tube containing PT100
3 - Platinum ring
4 - PTFE Junction
A special model (8350.3) is compatible with all the standard accessories and is designed for use in
hydrofluoric acid.
8350.4 8350.5 8351
Application Measuring pH at high
temperatures
Measuring pH in a
waste water
environment
Measuring redox
Junction type Flat, PTFE Flat, PTFE Flat, PTFE
Body material PPS CPVC PPS
Measurement range 0-14 pH 0-12 pH ± 1500 mV
Maximum temperature (°C) 110 80 110
Maximum pressure 10 bar 10 bar 10 bar
Drift (/ week) < 3 mV < 3 mV < 3 mV
Offset @ pH7 ± 20 mV ± 20 mV N/A
Slope 56-61 mV/pH 56-61 mV/pH N/A
Reference impedance @ 25°C < 50 Kohm < 50 Kohm < 50 Kohm
Glass impedance @ 25°C 150-500 Mohm 150-500 Mohm Platinum