Hach Polymetron Series - User Manual

Hach Polymetron Series
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
DOC023.98.80083
Polymetron Conductivity
Sensors
02/2017, Edition 3
User Manual
Bedienungsanleitung
Manuale dell'utente
Manuel d'utilisation
Manual del usuario
Manual do utilizador
Uživatelská příručka
Brugsanvisning
Gebruikershandleiding
Instrukcja obsługi
Bruksanvisning
Käyttäjän käsikirja
Ръководство за потребителя
Használati útmutató
Manual de utilizare
Naudotojo vadovas
Руководство пользователя
Kullanım Kılavuzu
Návod na obsluhu
Navodila za uporabo
Korisnički priručnik
Εγχειρίδιο Λειτουργίας
Kasutusjuhend
 

Table of Contents

Questions and Answers

  • A
    Anne MannAug 18, 2025
    What to do if my Hach Accessories say 'CAL OVERDUE'?
    • D
      Donna ChungAug 18, 2025
      If your Hach Accessories display a 'CAL OVERDUE' message, you should calibrate the sensor.
  • D
    davidsavageAug 20, 2025
    What to do if my Hach Accessories say 'NOT CALIBRATED'?
    • L
      Lisa WoodAug 20, 2025
      If the Hach Accessories show 'NOT CALIBRATED', you should calibrate the sensor.
  • M
    michelenunezAug 23, 2025
    What does 'REPLACE SENSOR' mean on my Hach Accessories, and what should I do?
    • M
      Michelle CarterAug 23, 2025
      If the Hach Accessories display 'REPLACE SENSOR' because the sensor has been in operation for more than 365 days, calibrate the sensor with a reference solution and reset the sensor days. If the calibration fails, call technical support.
  • S
    Samuel GonzalezAug 26, 2025
    What to do if my Hach Accessories displays 'ADC FAILURE'?
    • L
      Lisa MartinezAug 26, 2025
      If the Hach Accessories show 'ADC FAILURE' because the analog to digital conversion failed, ensure that the sensor module is fully inserted into the controller connector. If the issue persists, replace the sensor module.
  • N
    Nicholas NovakAug 28, 2025
    Why is my Hach Accessories showing 'SENS OUT RANGE'?
    • K
      Kelly HarrisAug 28, 2025
      The Hach Accessories display 'SENS OUT RANGE' because the sensor signal is outside of the accepted limits for the cell constant that is used. To resolve this, make sure that the sensor is configured for the correct cell constant.
  • R
    Ricardo DuncanAug 30, 2025
    What to do if my Hach Accessories show 'MEAS TOO HIGH'?
    • A
      Ashley RichardsonAug 30, 2025
      If the Hach Accessories show 'MEAS TOO HIGH' because the measured value is > 2,000,000 µS/cm, 1,000,000 ppm or 20,000 ppt, make sure that the sensor is configured for the correct cell constant.
  • A
    Alan NelsonSep 2, 2025
    What to do if my Hach Accessories show 'MEAS TOO LOW'?
    • R
      Rebecca ParrishSep 2, 2025
      If the Hach Accessories display 'MEAS TOO LOW' because the measured value is < 0 µS/cm, 0 ppm or 0 ppt, make sure that the sensor is configured for the correct cell constant.
  • D
    delgadodanielSep 4, 2025
    What does it mean when my Hach Accessories say 'OUTPUTS ON HOLD'?
    • S
      smithjasonSep 4, 2025
      If the Hach Accessories display 'OUTPUTS ON HOLD', the outputs will become active after the selected time period.
  • B
    bryan55Sep 6, 2025
    What does 'ZERO TOO HIGH' mean on my Hach Accessories?
    • L
      lisaroweSep 7, 2025
      If the Hach Accessories display 'ZERO TOO HIGH' because the zero calibration value is > 500,000 counts, make sure that the sensor is held in air during zero calibration and is not located near radio frequency or electromagnetic interference. Also, ensure that the cable is shielded by metal conduit.
  • B
    Bridget BlakeSep 9, 2025
    Why is my Hach Accessories showing 'TEMP TOO HIGH'?
    • L
      Lisa GreerSep 9, 2025
      The Hach Accessories display 'TEMP TOO HIGH' because the measured temperature is > 130 °C. To resolve this, make sure that the sensor is configured for the correct temperature element.

Summary

Operation

Configure the Sensor

Steps to enter sensor identification and adjust data handling options.

Calibrate the Sensor

Zero Calibration Procedure

Steps to define the unique zero point of the conductivity sensor before calibration.

Calibration with a Reference Solution

Adjusts sensor reading to match a reference solution value.

Calibration with the Process Sample

Calibrates the sensor using a sample from the actual process.

Troubleshooting

Test the Conductivity Sensor

Using an ohmmeter to test sensor resistance for calibration failures.

Error List

Displays and explains error codes and their resolutions for sensor issues.

Hach Polymetron Series Specifications

General IconGeneral
BrandHach
ModelPolymetron Series
CategoryAccessories
LanguageEnglish