EasyManua.ls Logo

Hach polymetron 9240 - Page 4

Hach polymetron 9240
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specification Details
Over voltage
category
2 (according to standard EN 61010-1)
Pollution degree 2 (according to standard CEI 664)
Altitude < 2000 m
Measurement
category
Cat II, Class 1 (overvoltage < 1500V)
Maximum panel
dimension (H x L x
D)
850 x 450 x 252.5mm [33.46 x 17.71 x 9.94in]
Inlet Simple fittings for 6 mm O.D. tubing or ¼" O.D. in PE-low density. ¼" OD in PHED-PTFE-
SS as option
Outlet Barbed stem for 12 mm (½" I.D.) hose
Protection rate Transmitter: IP65 (NEMA 4)
Panel: IP50 (dust protection)
Optional Enclosure: IP54 (splash water proof)
Instrument is designed to avoid DIPA vapor inside the enclosure. All DIPA vapor is
collected and sent to the instrument drain
Cell PMMA - compact (minimum tubing)
Flame rate Conform UL
Maximum weight 15 - 30 Kg
Mains power
supply
100 - 240 VAC, 50-60 Hz, ± 10%, automatic switching
Max. consumption 80 VA
Fuse 5 x 20 cartridge T2AL-250V following CEI127
Display Curve trend, Last Cal Date, Historical, Concentration, Temperature, Potential
Analog outputs Number: 6
4-20 or 0-20 mA (650 ohms)
Linear / Dual / Logarithm
Event indication
Relays 4 x Relay (conc)
1 x Warning
1 x System
Logic input Active / Inactive channels
Remote AutoCal
EMC requirements EN61326-1: EMC Directive
Note: This is a Class A product. In a domestic environment this product may cause radio interference in
which case the user may be required to take adequate measures.
N O T I C E
Measurement variations of less than 5% of the full range can occur if the instrument is
subject to a strong electromagnetic field.