Troubleshooting
Refer to the following table for common problem messages or
symptoms, possible causes and corrective actions.
Table 2 Calibration warnings and errors
Error/Warning Solution
Asymmetry >
58 mV
Slope out of range (accepted values ± 58 mV).
Repeat with .
Connect a new probe.
Buffer or electrode
in poor conditions.
Repeat with .
Examine the probe: Clean the probe (refer to Clean the
probe on page 9 for more information); make sure that
there are no air bubbles in the membrane. Shake the
probe like a thermometer; Connect a different probe to
verify if problem is with probe or meter.
Examine the buffer solution: Make sure that the buffer
used matches the buffer specified in configuration; make
sure of the temperature specification in configuration; use
new buffer solution.
Sens. (a) < 70%
BUFFER NOT
RECOGNIZED
UNSTABLE
READING
Time t> 100 s
Repeat with .
Examine the probe: Clean the probe (refer to Clean the
probe on page 9 for more information); make sure that
there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.
Make sure that the membrane and the diaphragm are
properly immersed in the sample.
Table 2 Calibration warnings and errors (continued)
Error/Warning Solution
Electrode in poor
conditions.
Examine the probe: Clean the probe (refer to Clean the
probe on page 9 for more information); make sure that
there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.
Check the
electrode
SAME BUFFERS
Repeat with .
Examine the probe: Clean the probe (refer to Clean the
probe on page 9 for more information); make sure that
there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.
Examine the buffer solution: Use new buffer solution.
Table 3 Measurement warnings and errors
Error/Warning Solution
pH 12.78 19°C
Repeat with .
pH out of range Examine the probe: Clean the probe (refer to Clean the
probe on page 9 for more information); make sure that
there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.
Temp out of range °C Examine the temperature sensor.
Connect a different probe to verify if problem is with
probe or meter.
Time > 150 s Make sure that the membrane and the diaphragm are
properly immersed in the sample.
Examine the temperature.
Examine the probe: Clean the probe (refer to Clean the
probe on page 9 for more information); make sure that
there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.
12 English