Figure 9 Battery replacement
1 Batteries 2 Release tab 3 Battery cover
Troubleshooting
Refer
to the following table for common problem messages or symptoms,
possible causes and corrective actions.
Error/
Warning
Description Solution
Measurement out
of range.
Insert the probe into the appropriate standard
and read again.
Examine the probe:
Clean the probe and the membrane.
Make sure that there are no air bubbles in
the membrane.
Disconnect the probe and then connect
the probe again.
Connect a different probe to verify if
problem is with probe or meter.
Battery is low. Insert new batteries.
E1 Unstable reading
during a stability
measurement or
during calibration.
Make sure that the probe is properly
immersed in the sample.
12 English