Chapter 6, Troubleshooting 6-33
Warning!
It is imperative that the operator record the codes listed if safety
system fault detection occurs. This information will be needed to
identify the reason for the safety system fault detection. It is also
imperative that the codes are properly labeled as COMMAND,
CHANNEL 1 FAULT, CHANNEL 2 FAULT, CHANNEL 1
CRITICAL, and CHANNEL 2 CRITICAL.
The command register defines the mode of operation the machine was in
when the fault occurred by displaying one of these hexadecimal codes.
Table 6-1, Error Codes of the Command Register
Code Critical Fault Description
000000 Start Test Phase Conducting system start-up tests.
000700 Stop PCS2 is in Ready mode or STOP or the
PUMP START/STOP key has been
pressed.
001C00 Start Application PCS2 is transitioning from the system
Phase start-up to the application mode.
000B00 Draw The CPU executes a Draw command.
001500 Return The CPU executes a Return command.
000D00 Pause The PCS2 is transitioning from state to
state.
001A00 Auto Prime The CPU executes a Prime command.
The fault register defines the actual fault that was detected. Each fault has
a hexadecimal code associated with it. If only one fault occurs, then one
of the codes listed in the following table will be displayed. If two or more
faults occur simultaneously, the actual code displayed is the sum of the
two or more codes.
!
Command
Register
Fault Register