Chapter 2, Machine Part Replacement Reference 2-25
Line
Conditioner
1. Remove the rear panel. (See page 2-2.)
2. Unplug the brown and blue AC power lines from the line conditioner
top & bottom.
3. Using a 11/32" wrench, remove the (2) #8-32 hex nuts securing the
line conditioner to the cabinet, pull the line conditioner away from the
cabinet.