2-26 PCS2 Service Manual
TTaabbllee 22-11,, PPCCSS22 CCoommppoonneenntt CCaalliibbrraattiioonn aanndd DDiiaaggnnoossttiicc MMaattrriixx
CALIBRATE
Replaced Part A/D: Cuff A/D: Weigher A/D: DPM Centrifuge Safety System Bowl Optics Line Sensor Power Supply
Processor PCB
Safety PCB
Driver PCB
Centrifuge Controller PCB
Front Panel Distribution PCB
Load Cell
DPM
Compressor
Centrifuge Assy
Photoelectric Assy X
Line Sensor
Bowl Optics Assy
Power Supply
DIAGNOSTIC CHECK
Replaced Part Pumps Valves Weigher Pressure Centrifuge Display Keyboard Transducers Air Detectors
Pump Motor (Blood or AC)
Pinch Valves
(Blood, Plasma, Saline)
Top Deck Distribution PCB
Membrane Panel Assy
Control Panel Distribution Cable
Centrifuge Distribution PCB
Air Detector
(BLAD, ACAD, DLAD 1 or 2)
Fluid Sensor Assy
Cover Switch Assy
Note: Shaded boxes indicate that a Calibration or Diagnostic Check of the appropriate part needs to be performed.