The PCS2 is designed to require minimal maintenance. The maintenance
the operator must perform consists primarily of cleaning the PCS2. A
record should be kept of the date and type of maintenance performed.
Haemonetics also recommends that a full preventive maintenance checkup
be performed annually by a Haemonetics Service Representative to assure
maximum machine performance.
As with any precision instrument, the PCS2 must be cleaned periodically.
The frequency and type of cleaning will depend on the number of
procedures performed with the PCS2. Following are instructions for
cleaning the various components of the PCS2.
Routine Cleaning Schedule
The following schedule shows the minimum recommended frequency for
cleaning the PCS2.
Component Source Plasma Blood Bank
Clean all exterior surfaces daily daily
Clean the donor pressure monitor (DPM) daily daily
Clean air detectors weekly monthly
Clean line sensor weekly monthly
Clean centrifuge well and chuck weekly monthly
Clean fluid sensor weekly monthly
Clean bowl optics lens weekly monthly
Clean pump rotors and wells monthly monthly
Clean air filters monthly monthly
Lubricate centrifuge O-ring monthly monthly
Chapter 8
Cleaning and Maintenance
Overview
Cleaning