6
Table 1: Delivery Time Information
Caution:
• Filling the pump less than the labeled volume results in faster ow rate.
• Filling the pump greater than the labeled volume results in slower ow rate.
Models
E050500 E100500 E101000 E102000 E251750 E252500 E401000 E402000 E502500
SE050500 SE100500 SE101000 SE102000
Labeled Fill Volume (ml) 50 100 100 100 250 250 400 400 500
Labeled Flow Rate (ml/hr) 50 50 100 200 175 250 100 200 250
Minimum Fill Volume (ml) 28 60 65 60 200 250 200 200 300
Maximum Fill Volume (ml) 65 125 125 125 335 335 500 550 550
Retained Volume (ml) ≤3 ≤3 ≤3 ≤3 ≤8 ≤8 ≤10 ≤10 ≤10
Time to Reach Room Temperature
Hours from refrigerator 7 8 8 8 12 12 15 15 18
Hours from freezer 12 16 16 16 25 25 29 29 30
Approximate Delivery Time Fill Volume (ml)
00:15 h 60
00:30 h 28 65 100
00:40 h 125
00:45 h 200
01:00 h 50 60 100 200 250 250 300
01:15 h 65 125
01:20 h 250 335 300
01:30 h 80 400
01:35 h 200
01:40 h 350
01:45 h 300
02:00 h 100 335 250 400 500
02:15 h 450 550
02:30 h 120
02:35 h 300
02:40 h 125 500
03:00 h 350
04:00 h 400
04:45 h 450
05:15 h 500