PRIMING THE PUMP
Hi control mode on the BL100
Set
point
Set
point
Pump statusPump status
eresis Hyster
Measured
Measur
RUNNING RUNNING
NOT RUNNING NOT RUNNING
Pump running continuously
at the congured ow rate
Pump running continuously
at the congured ow rate
Lo control mode on the BL100
Proportional Control
With Proportional control enabled in SETUP, the dosing time depends on the difference between pH
(or ORP) value and assigned setpoint.
Set
point
Set
point
Regulator
output
Regulator
output
100% 100%
Band
time
proportional mode
with 0.25L/h
time
proportional mode
with 0.25L/h
Band
speed proportional
0.25L/h
ow rate
SETUP
0.3÷3.5L/h
Flow rate
Set value
Measured
value
Measured
value
Flow rate
Set value
speed proportional
0.25L/h
ow rate
SETUP
3.5÷0.3L/h
0.25L/h
90%
50%15%
0.25L/h
90%
50% 15%
with Hi Control Mode and Band as illustrated
with Lo Control Mode and Band as illustrated
An overview of pH proportional control is presented in the graph below.
0.1
The same principle applies to ORP control.
8.2. PRIMING THE PUMP
• From measurement mode, long press the arrow keys together to manually activate the pump.
• To exit, release the two keys.
• When there are no errors or active alarm the LCD backlight is light green.