Notes: • Buffers (standards) displayed in video inverse mode are
from previous calibrations (pH and ISE range only).
• The pH custom buffers are marked with an “*” on the
right side of the buffer value.
LAST RELATIVE mV CALIBRATION DATA
Last Relative mV calibration data is stored automatically after a successful
calibration.
To view the Relative mV calibration data, press GLP key while in Relative
mV measurement mode.
The instrument will display the Relative mV GLP information: calibration
date, time and offset.
LAST ISE CALIBRATION DATA
Last ISE calibration data is stored automatically after a successful
calibration.
To view the ISE calibration data, press GLP function key while in ISE
measurement mode. If GLP function key is not displayed press MENU
key.
The instrument will display the ISE calibration information: calibration
date, time, slope, calibration status and electrode type.
LAST EC CALIBRATION DATA
The last pH calibration data is stored automatically after a successful
calibration.
To view the EC calibration data, press GLP when the instrument is in the
EC measurement mode.
The instrument will display a lot of data including calibration standards,
offset, time and date, cell constant, etc.
LAST NaCl % CALIBRATION DATA
Last NaCl calibration data is stored automatically after a successful
calibration.
To view the NaCl calibration data, press GLP key while in NaCl measurement
mode.
The instrument will display the NaCl information: calibration date, time
and salinity factor.
Notes: • Press ESC at any moment and the instrument will return to
measurement mode.
• If calibration has not been performed, the instrument
displays “No user calibration” message.