5150
If Standard was pressed.
Use ARROW keys to focus on the desired electrode.
Press Accept to confirm setting or ESC to exit.
Press View to see probe parameters.
ISE unit
Focus on the
ISE Unit
item.
Press Modify.
Use ARROW key to select unit.
Press Accept to confirm selection or ESC to exit.
Notes: • If the unit is changed or “User” is selected a warning
message will be displayed to alert that the ISE range must
be calibrated.
• If a new probe was selected or custom probe parameter are
changed, the ISE range must be calibrated.
Focus on
Charge/Slope
item.
Use ARROW keys in order to select the desired combination.
If “None/-59.16“ is selected the slope of the probe can be changed by
pressing Modify key.
Press Modify.
Use ARROW keys to change the slope.
Press Accept to confirm or ESC to exit.
Focus on
Molar Weight
.
Press Modify in order to change molar weight.
Use ARROW keys to change the value. Press Accept to confirm or ESC to
exit.