EasyManuals Logo

Hanna Instruments HI 83214 User Manual

Hanna Instruments HI 83214
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
25
Remove the cap and add the content of one packet of HI 93701-0 Free
Chlorine Reagent.
Replace the cap and shake gently to mix for about 20 seconds. This is the
sample.
Place the vial into the holder and push it completely down.
Press Timer and the display will show the countdown prior to the measurement and the message
Reaction Time”. Alternatively, wait for 1 minute and press Read. The instrument will perform the
reading.
The instrument displays concentration in mg/L of free chlorine.
INTERFERENCES
Interferences may be caused by:
Bromine (Br
2
)
Iodine (I
2
)
Oxidized forms of Chromium and Manganese
Ozone (O
3
)
Alkalinity above 250 mg/L CaCO
3
or acidity above 150 mg/L CaCO
3
will not reliably develop the full amount
of color or it may rapidly fade. To resolve this, neutralize the sample with diluted HCl or NaOH.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately
2 minutes after adding the powder reagent.
Chlorine, Free

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 83214 and is the answer not in the manual?

Hanna Instruments HI 83214 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 83214
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals