Press Start and the display will show the countdown prior to the measurement and the ”Reaction
Time” message. Alternatively, wait for 5 minutes and press Zero.
The instrument will perform a zero sequence and after a few seconds the display will show
-0.0-”. Now the meter is zeroed and ready for measurement.
Remove the blank vial and place the sample vial into the holder and push it completely down.
Press Read. The instrument will perform the reading. The instrument displays the concentration in
mg/L of total nitrogen (N).
Press s or t to convert the result in mg/L of ammonia (NH
3
) or mg/L of nitrate (NO
3
-
).
The method detects all organic and inorganic forms of nitrogen present in the sample.
INTERFERENCES
Interference may be caused by:
Bromide (Br
-
) above 240 mg/L
Chloride (Cl
-
) above 3000 mg/L
Chromium (Cr
3+
) above 0.5 mg/L
Nitrogen, Total HR