EasyManuals Logo

Hanna Instruments HI 83214 User Manual

Hanna Instruments HI 83214
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
44
Choose a homogeneous sample. Samples containing settleable solids need to be homogenized with a
blender.
Preheat the Hanna Reactor HI 839800 to 150 °C (302 °F). For correct use of the reactor follow Reactor
Instruction Manual.
Use of the optional HI 740217 safety shield is strongly recommended.
DO NOT USE AN OVEN OR MICROWAVE samples may leak and generate a corrosive and possibly explosive
atmosphere.
Remove the caps from two Reagent Vials.
Add exactly 2.0 mL of sample to one vial (sample vial), and 2.0 mL of deionized water to the other
vial (blank vial), while keeping the vials at a 45-degree angle. Replace the cap tightly and mix by
inverting each vial a couple of times.
Warning: The vials will become very hot during mixing, be careful
when handling them.
Insert the vials into the reactor and heat them for 2 hours at 150 °C.
At the end of the digestion period switch off the reactor. Wait for twenty
minutes to allow the vials to cool to about 120 °C.
Invert each vial several times while still warm, then place them in a test
tube rack.
Warning: The vials are still hot, be careful when handling them.
Leave the vials in the tube rack to cool to room temperature. Do not shake
or invert them, the samples may become turbid.
Blank
Sample
Blank
Sample
Blank
D.I. Water
Sample
Sample
COD MR
white label

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 83214 and is the answer not in the manual?

Hanna Instruments HI 83214 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 83214
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals