Page #60 background imageLoading...
Page #60 background image
60
Blank
Sample
Blank
D.I. Water
Sample
Sample
Phosphorus, Reactive HR
Replace the cap and mix by inverting each vial a couple of times.
Place the blank vial into the holder and push it completely down.
Press Timer and the display will show the countdown prior to the measurement and the “Reaction
Time” message. Alternatively, wait for 7 minutes and press Zero.
The instrument will perform a zero sequence and after a few seconds the display will show “-0.0-”. Now
the meter is zeroed and ready for measurement.
Remove the blank vial.
Place the sample vial into the holder and push it completely down.
Press Read. The instrument will perform the reading. The instrument displays concentration in mg/L of
phosphorus (P).
Press s or t to convert the result in mg/L of phosphate (PO
4
3
-
) or mg/L of P
2
O
5
.

Questions and Answers: