EasyManuals Logo

Hanna Instruments HI 83214 User Manual

Hanna Instruments HI 83214
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
64
Blank
Sample
Blank
Sample
Blank
Sample
Blank
Sample
Phosphorus, Total HR
Select the
Phosphorus,
TT
TT
Total High Range
method following one of the procedures described in the
“Method Selection” section (see page 15).
Remove the cap from the vials and add exactly 2.0 mL of Sodium Hydroxide (NaOH) Solution 1,54 N
to each vial, while keeping the vials at a 45-degree angle. Replace the cap tightly and mix by
inverting the vials a couple of times.
Remove the cap from the vials and add exactly 0.5 mL of HI 93763B-0 Molybdovanadate Reagent
to each vial, while keeping the vial at a 45-degree angle. Replace the cap tightly and mix by
inverting the vials a couple of times.
Place the blank vial into the holder and push it completely down.
Press
Timer and the display will show the countdown prior to the measurement and the “Reaction
Time” message. Alternatively, wait for 7 minutes and press Zero.
The instrument will perform a zero sequence and after a few seconds the display will show “-0.0-”. Now
the meter is zeroed and ready for measurement.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 83214 and is the answer not in the manual?

Hanna Instruments HI 83214 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 83214
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals