Press and hold down Temp to display the
temperature.
Adjust the TEMPERATURE knob to display 20°C.
Release Temp key to display conductivity measurement.
Select 19.99 mS/cm (HI 8633)
or 1999 mg/L (HI 8734) range
by pressing the appropriate
range key.
Follow LCD calibration indication. Adjust the calibration knob until
the display shows for HI 8633 the conductivity reading at 25°C
(see the conductivity vs. temperature table), e.g. @ 25°C,
12880 µS/cm = 12.88 mS/cm, or for HI 8734 the TDS
reading at 25°C (see the TDS vs. temperature chart), e.g.
@25°C, 1382 mg/L.
All subsequent measurements will be compensated to 25°C (77°F).
If you prefer to standardize the temperature compensation to
20°C (68°F) rather than 25°C (77°F), leave the TEMPERATURE
knob at 18°C (if the temperature of the solution is 18°C), adjust
the knob to read “11.67 mS” (see the conductivity vs. tempera-
ture chart) or “1251 mg/L” (see the TDS vs. temperature chart). All
subsequent measurements will be compensated to 20°C.
The calibration is now complete and the instrument is ready for
use.
www. .com
information@itm.com1.800.561.8187

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 8733 and is the answer not in the manual?

Hanna Instruments HI 8733 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 8733
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals