11
The instrument should be recalibrated at least once a month, or
when the probe is changed.
Note: For more accurate results, it is advisable to use a calibration
solution close to the measurement range. See the “Accessories”
section for a wide selection of conductivity solutions.
PROCEDURE FOR HI 8733
Pour sufficient quantity of a conductivity
calibration solution (e.g. HI 7030/
HI 8030) into a beaker to cover the
holes on the probe. If possible, use
plastic beakers to minimize any EMC
interference.
Immerse the conductivity probe in the
solution, making sure that holes are
completely submerged.
Wait for a couple of minutes for thermal
equilibrium to be reached.
Tap the probe on the bottom, then
shake it while rotating to make sure no
air bubbles remain trapped in the
sleeve.
Switch the instrument on by pressing
On/Off.
Set the temperature coefficient knob to
2% to compensate for the temperature
effect of average solutions (to determine
exact value for a particular solution,
see page 15).
Select 19.99 mS/cm range by pressing
the appropriate range key.
www. .com
information@itm.com1.800.561.8187

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 8733 and is the answer not in the manual?

Hanna Instruments HI 8733 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 8733
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals