To delete last calibration, follow next steps:
Enter the GLP feature by pressing RCL/GLP for 3
seconds.
The date of the last calibration will be displayed
on the LCD.
If no calibration was performed, the “F.CAL”
message appears on the LCD and the instrument
returns to measurement mode.
Press READ/TIMER
to see the information
related to calibration. The last panel is the one
with delete calibration.
Press SETUP/DEL to delete the current calibration.
The instrument will restore the factory calibration
and will automatically return to measurement
mode.
CALIBRATION DELETION
HI 93414 is delivered factory calibrated. It is possible to restore factory calibration at any time if the
user calibration do not work as expected.
Note: Deleting the user calibration for one range will not affect the other ranges.
CALIBRATION ERROR MESSAGES
The calibration is successfully performed if the
CAL CHECK™ readings is in certain limits.
If the CAL CHECK™ standard value is too high,
the display will show “-HI-” on the primary
display and “Err” on the secondary display. If
this message appears, check if the correct cuvet
was used.
If the CAL CHECK™ standard value is too low,
the display will show “-LO-” on the primary
display and “Err” on the secondary display. If
this message appears, check if the correct cuvet
was used.