APPENDIX A - PROBE MAINTENANCE
The HI 7698292 probe maintenance kit includes HI 7042S (electrolyte solution
for D.O. sensor), spare membranes with o-rings for D.O. sensor, a small brush
for cleaning EC , o-rings for sensor connectors and a syringe with grease to
lubricate these o-rings.
General Maintenance
Inspect all sensor connectors for corrosion and replace sensors if necessary.
Inspect sensor o-rings for nicks or other damage and replace sensor if necessary.
Lubricate only with grease from kit.
Use only the supplied grease as some lubricants can cause the
o-rings to expand or affect the turbidity calibration standards.
After prolonged storage or cleaning, calibration of the sensors is required.
After use rinse the probe with tap water and dry it. The pH electrode bulb must
be kept moist. Dry the D.O. and EC sensors.
Check GLP data under “Status” to ensure the sensor is still functioning properly.
pH and pH/ORP Sensor Maintenance
Remove the sensor protective cap. Do not be alarmed if any salt deposits are
present. This is normal with pH/ORP electrodes and they will disappear when
rinsed with water.
Shake down the sensor as you would do with a clinical thermometer to eliminate
any air bubbles inside the glass bulb.
If the bulb and/or junction are dry, soak the electrode in HI 70300 storage
solution for at least one hour.
To ensure a quick response time, the glass bulb and the junction should be kept
moist and not allowed to dry. Store the sensor with a few drops of HI 70300
storage solution or pH 4.01 buffer in the protective cap. Tap water may also be
used for a very short period (few days).
Never use distilled or deionized water to store pH sensors
Inspect the sensor for scratches or cracks. If any are present, replace the sensor.
Cleaning procedure: clean the sensor frequently by soaking it for 1 minute in
HI 70670 or HI 70671 cleaning solution. After cleaning soak the sensor in
HI 70300 storage solution before taking measurements.
‘’Warning x’’: Any other warning that appears at power-on is identified using
a numeric code. Restart the meter. If the problem persists, contact the
HANNA service center. Some meter/probe features can be accessed but
with no guarantee.
‘’Errors x’’: Any critical errors that appear are identified using a numeric code,
and the meter is automatically switched off. Contact the HANNA service
center.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 98194 and is the answer not in the manual?

Hanna Instruments HI 98194 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 98194
CategoryMultimeter
LanguageEnglish

Related product manuals