EXPIRED CALIBRATION ALARM
Press CAL when the calibration time‑out is displayed. Calibration time‑out
(“OFF” or “1” to “7” days) and “CFM” will start blinking.
Press the ARROW keys to change the calibration time‑out value.
Press CFM to save the modified calibration time‑out value.
Press CAL to escape without saving.
LOG INTERVAL
Press CAL when log interval is displayed. The log interval and “CFM
will start blinking (“LOG” for log on demand, “StAb” log on stability,
interval in seconds or minutes).
Press the ARROW keys to change the custom buffer value.
Press CFM to confirm the selection.
Press CAL to escape without saving.
FIRST CUSTOM BUFFER
Press CAL when “Cust. Buff. 1” is displayed. The custom buffer (disabled
“no” or “0” to “16” pH) and “CFM” will start blinking.
Press the ARROW keys to change the custom buffer value.
Press CFM to save the modified custom buffer value.
Press CAL to escape without saving.