PROCEDURE FOR HI8733
Pour sufficient quantity of a conductiv
ity calibration solution (e.g. HI7030/
HI8030) into a beaker to cover the
holes on the probe. If possible, use
plastic beakers to minimize any EMC
interference.
Immerse the conductivity probe in the
solution, making sure that holes are
completely submerged.
Wait for a couple of minutes for
thermal equilibrium to be reached.
Tap the probe on the bottom, then
shake it while rotating to make sure
no air bubbles remain trapped in the
sleeve.
Switch the instrument on by pressing
On/Off.
Set the temperature coefficient knob
to 2% to compensate for the tem
perature effect of average solutions (to
determine exact value for a particular
solution, see page 15).
Select 19.99 mS/cm range by pressing
the appropriate range key.
The calibration is now complete and the instrument is ready for
use.
The instrument should be recalibrated at least once a month, or when
the probe is changed.
Note: For more accurate results, it is advisable to use a calibration
solution close to the measurement range. See the “Accessories”
section for a wide selection of conductivity solutions.