EasyManuals Logo

Hanna Instruments HI8733 User Manual

Hanna Instruments HI8733
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
12
Follow LCD calibration indication. Adjust the calibration knob
until the display shows “12.88 mS” i.e. the conductivity reading
@ 25 °C.
All subsequent measurements will be compensated to 25 °C
(77 °F).
If you prefer to standardize the temperature compensation to
20 °C (68 °F) rather than 25 °C (77 °F), adjust the knob to read
11.67 mS” (see the conductivity vs. temperature chart on page
13). All subsequent measurements will be compensated to 20 °C.
The calibration is now complete and the instrument is ready for
use.
The instrument should be recalibrated at least once a month, or when
the probe is changed.
Note: For more accurate results, it is advisable to use a calibration
solution close to the range to be measured. See the “Accessories”
section for a wide selection of conductivity solutions.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI8733 and is the answer not in the manual?

Hanna Instruments HI8733 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI8733
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals