SETUP
Note: If enabled “CAL DUE” warning will be displayed, the set number of days after
calibration is over passed.
First Point Mode
Highlight First Point Mode.
Press the displayed functional key in order to change the option.
If the Replace option is selected, the slopes between current buffer and nearest lower and higher
buffers will be recalculated.
If the Offset option is selected, an electrode offset correction is performed keeping unchanged the
existing slopes.
Custom Buffers
Highlight Custom Buffers.
Press Modify.
Press Delete to delete custom buffer value.
Press Add to add a new buffer to the list (max 5).