32
SETUP
Press Modify to set custom buffer value.
Use / keys to change the value.
Press Accept to confirm custom buffer value or ESC to exit without saving.
View Calibration Points
Highlight View Calibration Points.
Press the displayed functional key to change option.
If option is enabled the calibration buffers corresponding to the last calibration are displayed in the
pH measurement screen.
Out of Calibration Range Warning
Highlight Out of Cal.Range Warning.
Press the displayed functional key in order to change option.
If enabled, the “Out Cal Range” message will be displayed if the pH reading is outside.
Approximately 1 pH unit from the lowest and highest pH buffer value used for calibration. For
single point pH 7.01 calibration the message will be displayed for pH values less than 4 or greater
than 10.