29
Measurement mode is one of the three main operating modes of HI9829 (along with logging mode
and setup mode).
During measurement mode HI9829 will simultaneously measure data for all enabled parameters.
• Use the numbers on the keyboard to select the number of parameters that are shown on the screen
at one time. The display will automatically resize the font.
• Press the and keys to scroll through the enabled parameters if they do not fit on one screen.
Note: A flashing measurement value indicates that the measurement is out of range.
A flashing measurement unit indicates that the user calibration has not been done and
is
needed for accurate readings.
• Press <Log> to enter the log menu. See Chapter 11 for details.
Press <Menu> to enter the main setup menu. The main menu accesses the parameter setup,
calibration, system setup, GPS and status options. See the following chapters for details.