36
The quick calibration method provides a quick single point calibration for pH, conductivity and dissolved
oxygen sensors. HI9828‑0 calibration solution is used for both pH and conductivity.
• Fill the calibration beaker 2/3 full with HI9828‑0 calibration solution.
• Slowly place the sensors into the solution and dislodge
bubbles that may adhere to the sensors.
• Screw the calibration beaker completely on the probe body.
Some solution may overflow.
• Wait a few minutes for the system to stabilize.
• From the “Calibration” menu select “Quick calibration”.
• A three item calibration menu will appear (pH, Conduc‑
tivity and Dissolved oxygen) and “pH” will start to blink
along with the “Not ready” message.
• When the pH signal is stable, the “Ready” message
appears. Press <Confirm> to store the calibration data.
• The “Storing” message will appear as the calibration
proceeds to the next sensor. A checkmark will appear in
the box next to “pH” to indicate a successful calibration.
Note: To bypass any of the calibrations press <Skip> to
move to the next sensor in the quick calibration menu.
If the pH sensor is not installed the message “pH
sensor not installed! Skip to conductivity calibration”
will appear.
• Following the pH calibration, “Conductivity” will start to blink along with the “Not ready” message.
• When the measurement is stable, “Ready” appears. Press <Confirm> to store the calibration
data and the “Storing” message will appear.
Note: If EC calibration is not required, skip to the D.O.
quick calibration by pressing the <Skip> softkey.
• The message “Empty the beaker.” will appear.
QUCIK CALIBRATION