The quick calibration method provides a quick single point calibration for pH, conductivity and dissolved
oxygen sensors. HI9828‑0 calibration solution is used for both pH and conductivity.
• Fill the calibration beaker 2/3 full with HI9828‑0 calibration solution.
Slowly place the sensors into the solution and dislodge
bubbles that may adhere to the sensors.
Screw the calibration beaker completely on the probe body.
Some solution may overflow.
• Wait a few minutes for the system to stabilize.
• From the “Calibration” menu select “Quick calibration”.
• A three item calibration menu will appear (pH, Conduc
tivity and Dissolved oxygen) and “pH” will start to blink
along with the “Not ready” message.
When the pH signal is stable, the “Ready” message
appears. Press <Confirm> to store the calibration data.
The “Storing” message will appear as the calibration
proceeds to the next sensor. A checkmark will appear in
the box next to “pH” to indicate a successful calibration.
Note: To bypass any of the calibrations press <Skip> to
move to the next sensor in the quick calibration menu.
If the pH sensor is not installed the message “pH
sensor not installed! Skip to conductivity calibration”
will appear.
Following the pH calibration, “Conductivity” will start to blink along with the “Not ready” message.
When the measurement is stable, “Ready” appears. Press <Confirm> to store the calibration
data and the “Storing” message will appear.
Note: If EC calibration is not required, skip to the D.O.
quick calibration by pressing the <Skip> softkey.
• The message “Empty the beaker.” will appear.
QUCIK CALIBRATION