EasyManua.ls Logo

Hanna Instruments HI9829 - Page 38

Hanna Instruments HI9829
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
Procedure
The measured pH value is displayed, along with the temperature and the buffer value on the second
level.
If necessary, press the <Cal point> softkey and use the / keys to select the correct buffer.
Immerse the sensors in the first buffer rinse solution and stir gently.
Immerse the pH sensor and temperature probe into
the selected buffer and stir gently. The temperature, pH
buffer value and the “Not ready” message are displayed.
Once the reading has stabilized the countdown timer will
count down until the display shows the “Ready” message.
• Press <Confirm> to accept the calibration point.
After the calibration point is confirmed, to avoid cross‑
contamination immerse the sensors in the next calibration
buffer rinse solution and stir gently.
Press <Cal Point> to select the next buffer (if neces
sary), and repeat the calibration procedure outlined above
with the second and third buffers.
Note: The calibration procedure can be terminated after a single or 2 point calibration by press-
ing <ESC>. The message “Storing” followed by “Calibration completed” will be displayed.
• Press <OK> to return to the Calibration menu.
• Press <Measure> to return to the measurement screen.
Custom buffer calibration
• The HI9829 permits a single custom buffer to be used for
pH calibration. This can be used along with standard buffers
as part of a 2 or 3 point calibration or as a single point.
To select this option first press <Cal. point> and then
<Custom> while the meter is waiting for stable reading.
A text box window will appear. Use the keypad to enter
the value of the buffer at the current temperature. The valid range for custom a buffer is from
0.00 to 14.00 pH.
pH CALIBRATION

Related product manuals