47
Salinity
The measurement of salinity is based on the Practical Salinity Scale which uses the EC measurement.
If the user has a standard with known PSU value it may be used to calibrate the conductivity sensor.
Select “Salinity” from the “Conductivity calibration” menu.
Use the keypad to enter the known salinity value of the
calibration solution. Press <Accept> to confirm.
• Fill the calibration beaker with salinity standard of known
value.
• Pour additional standard into a second beaker to be used to rinse the sensor.
Immerse the sensor into the rinse beaker and raise and lower the beaker to ensure that the EC
sensor channels are filled with fresh standard.
Place the calibration beaker with standard over the EC sensor and dislodge any trapped gas bubbles.
Screw the beaker into place.
Wait for the reading to stabilize. The stability timer will count down and <Confirm> will appear.
• Note the temperature and adjust the salinity value if needed.
• Press <Confirm> to save the calibration.
After confirmation, the following messages are displayed: “Storing” and “Calibration com
pleted”.
• Press <OK> to return to the “Calibration” menu.
• Press ESC twice to return to the main menu.
• Press <Measure> to return to the measurement screen.
• Note the temperature and adjust the conductivity value if needed.
• Press <Confirm> to save the calibration.
• After confirmation, the following messages are displayed: “Storing” and “Calibration completed”.
• Press <OK> to return to the “Calibration” menu.
• Press ESC twice to return to the main menu.
• Press <Measure> to return to the measurement screen.
CONDUCTIVITY CALIBRATION