The probe is factory calibrated for temperature readings. The user can perform a single point
temperature calibration or restore factory calibration. This procedure requires a reference temperature
measuring instrument.
• Select “Temperature” from the “Calibration” menu.
• Select “Calibrate temperature”.
Insert the probe in an isothermal bath with reference instrument and allow time for the probe to
come to thermal equilibrum.
• Use the keypad to enter the known temperature and then press <Accept> to confirm.
• The stability timer will count down and the message “Ready” and <Confirm> will be displayed.
• Press <Confirm> to store the calibration point.
• After confirmation, the following messages are displayed “Storing” and “Calibration completed”.
• Press <OK> to return to the “Calibration” menu.
• Press <Measure> to return to the measurement screen.
To restore the factory calibration, select the corresponding option in the “Temperature calib.” menu
and then press <Select>.