SCPI Programming Reference
HDG2000 Series Arbitrary Waveform Generator 81
SOURce<n>:HARMonic:ORDEr <value>|MINimum|MAXimum
SOURce<n>:HARMonic:ORDEr? [MINimum|MAXimum]
Set Harmonic order.
Parameter
Typical Return
2 to the highest frequency ÷fundamental
frequency, ,integer, Max 16, default 2
+8.000000000000000E+00
Set Harmonic order on CH1 to 8: SOURce1:HARMonic:ORDEr 8
SOURce<n>:HARMonic:TYPe EVEN|ODD|ALL|USER
SOURce<n>:HARMonic:TYPe?
Select Harmonic type even, odd or all.
Parameter
Typical Return
EVEN|ODD|ALL, default ALL
EVEN, ODD, or ALL
Set harmonic type on CH1 to odd: SOURce1:HARMonic:HARMonic:TYPe ODD
5.2.14 IEEE-488 Common Commands
*IDN? - Instrument identification
*RCL/*SAV - Recall/save instrument state
*RST - Reset instrument to factory defaults
*TRG - Trigger command, command is valid under sweep mode and burst mode
*TST? - Self-test
*IDN?
Identification Query. Returns instrument’s identification string.
Parameter
Typical Return
none
Return factory, model, series number andvesion
Return the instrument's identification string: *IDN?
*RST
Resets instrument to factory default state.
Parameter
Typical Return
none
none
Reset the instrument: *RST
Does not affect stored instrument states, stored arbitrary waveforms, or I/O settings; these are
stored in non-volatile memory.