Patient Testing
Obtain patient sample by fingerstick:
After cleaning the patient’s finger
with alcohol swab and letting dry
use a lancet to prick it.
Wipe away the first 2 or 3 drops of
blood with a lint free wipe (gauze).
Do not use cotton balls!
Apply light pressure towards the
fingertip until another drop of blood
appears. Do not squeeze the finger.
Hold the microcuvette with the
pointed end away from you.
Touch the pointed tip of the
microcuvette to the drop of the blood
droplet.
Allow the microcuvette to fill by
capillary action. Do not attempt to
refill.