Questions and Answers

  • B
    Brittany FloydNov 27, 2025
    Why am I getting unexpected patient or control results from my HemoCue 201 DM?
    • M
      Mary HayesNov 27, 2025
      Unexpected patient or control results from the HemoCue Docking Station may be due to: 1. Improper sampling technique. 2. Expired, faulty, or improperly stored cuvettes. Check the expiration date and storage conditions. 3. Contamination on the cuvette's optical eye. Use a new cuvette for measurement. 4. Air bubbles in the cuvette. Remeasure with a new cuvette. 5. Dirty optical parts. Clean the optical parts. 6. Incompletely filled cuvette. Use a new cuvette for measurement. 7. Delay in starting the measurement after filling the cuvette. Use a new cuvette and repeat the measurement. 8. The control has not been properly mixed and/or has not reached room temperature. Make sure that the control is properly mixed and that it has reached room temperature. 9. Control solution not compati...
  • C
    Catherine MckayAug 25, 2025
    How to fix HemoCue 201 DM Docking Station no transfer of data via USB?
    • K
      Kelly MillerAug 26, 2025
      To troubleshoot USB data transfer issues with your HemoCue Docking Station: a) Ensure the Analyzer is properly docked. b) Check cable connections at both the Docking Station and PC. c) Verify the Docking Station isn't a Primary Docking station. d) Confirm all Docking Stations are connected to the PC with a USB cable. e) Ensure the DMS Software and USB driver are correctly installed. f) Check the electrical connection; the LED on the docking station should display a stable or flashing green light.
  • M
    mclaughlinmitchellAug 26, 2025
    What to do if HemoCue Docking Station has no characters on the display?
    • A
      Ashley WestAug 26, 2025
      If your HemoCue Docking Station display shows no characters: 1. Check the power adapter connection to both the power supply and the Docking Station. 2. If the Analyzer is docked, ensure the green LED on the Docking Station is flashing. 3. If using battery power, recharge the battery using a power adapter or Docking Station. If the issue persists, the display may be faulty.
  • S
    Sara PerkinsAug 27, 2025
    Why is my HemoCue 201 DM Docking Station analyzer not charging?
    • A
      Angela HughesAug 27, 2025
      If your HemoCue analyzer is not charging, ensure that: a) The analyzer is properly docked. b) The green LED on the docking station flashes green when the analyzer is docked. If the problem continues, consider replacing the battery.
  • J
    Jason MartinezAug 30, 2025
    How to troubleshoot a malfunctioning scanner on my HemoCue Docking Station?
    • K
      Katherine ValdezAug 30, 2025
      If the scanner on your HemoCue Docking Station is malfunctioning: 1. Verify you are scanning the correct product barcode. 2. Check the product's expiration date. 3. Ensure the analyzer is 10–30 cm (4–12 inches) from the barcode. 4. If the barcode is indistinct, enter the information manually. 5. Clean the scanner glass. 6. Ensure the barcode is compatible with the scanner (see Technical Specifications).
  • M
    Meagan RiveraSep 23, 2025
    What causes error E08 on HemoCue 201 DM Docking Station?
    • R
      ryan97Sep 23, 2025
      Error E08 on the HemoCue Docking Station is caused by a light-blocking item in the Cuvette holder. Check that the Analyzer and Cuvettes are used according to the Instructions for Use.
  • H
    hernandezkristaSep 17, 2025
    What does HemoCue 201 DM error E23 mean?
    • K
      Kevin ClarkSep 17, 2025
      HemoCue Docking Station error E23 indicates that the Real Time Clock backup battery has been drained and needs to be replaced. Contact distributor.
  • A
    Ashley OrrSep 18, 2025
    What to do if HemoCue 201 DM Docking Station displays error E30?
    • G
      Gabriel SanchezSep 18, 2025
      To address error E30 on your HemoCue Docking Station, turn off the Analyzer and clean the optronic unit.
  • G
    gomeztheresaSep 8, 2025
    What does error E28 mean on my HemoCue Docking Station?
    • A
      Amy DavisSep 9, 2025
      Error E28 on the HemoCue Docking Station signifies that the Analyzer log memory is full, preventing the saving of new Error Codes and Log Notes. To address this, save test results by connecting the Analyzer to DMS Software or an OR via a Docking Station. With the necessary authority, you can delete Analyzer logs or set the Analyzer to 'Overwrite' when the QC test memory is full.
  • D
    Derrick FlemingSep 6, 2025
    What does error E26 mean on my HemoCue 201 DM Docking Station?
    • M
      Matthew RobinsonSep 6, 2025
      Error E26 on your HemoCue Docking Station means the Patient test memory is full, and no more patient data can be saved. To resolve this, save test results by placing the Analyzer in a Docking Station connected to DMS Software or OR. With supervisory authority, you can also delete stored Patient Tests or change Analyzer settings to 'Overwrite' when the memory is full.

Summary

Docking Station Overview

Placing the Analyzer in the Docking Station

Instructions on how to correctly insert and remove the analyzer from the docking station.

Docking Station General Information

Getting Started: Single Docking Station

Connecting to a PC or LAN

Provides instructions for connecting a single docking station to a PC or LAN.

Connecting Primary Docking Station to PC

Details the steps to connect a primary docking station to a PC via USB.

Connecting Primary Docking Station to LAN

Outlines the procedure for connecting a primary docking station to a LAN.

Connecting Secondary Docking Station to PC

Explains how to connect a secondary docking station to a PC via USB.

Power Source and Connections

Connecting the AC Adapter

Instructions for connecting the AC adapter to power the docking station.

Docking Station LED Indicators

Troubleshooting Guide

Troubleshooting Data Transfer Issues

Lists symptoms and actions for troubleshooting data transfer problems.

Troubleshooting USB Data Transfer

Addresses problems related to data transfer via USB connection.

Troubleshooting Analyzer Charging

Provides troubleshooting steps when the analyzer is not charging.