EasyManua.ls Logo

Heyl Testomat 808 - Further possible instrument errors; Response of a protective circuit; Malfunctioning;Repairing a defective instrument

Heyl Testomat 808
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error messages/Troubleshooting
36
Further possible instrument errors
Error
Possible causes
Remedies
Instrument not functioning,
even though it is switched
on
Fuses F1, F2 or F3 defective
Power switch defective
Ribbon cable to controller board
or
base circuit board is loose
Error at controller board or base
circuit board
Replace fuses
Replace power switch
Reconnect ribbon cable
Replace controller or base
circuit board
Response of a protective circuit
After a protective circuit (fuse) has been tripped, attempt to eliminate
the cause of malfunctioning (e.g. replace a defective valve) before
reactivating the protective circuit. Frequent tripping is always due to
an error which, in certain circumstances, may also cause damage to
the instrument.
Malfunctioning/Repairing a defective instrument
The repair of a defective instrument irrespective of the warranty
period - is only possible when the instrument is dismantled and
returned to us with a description of the error. Please also inform us
about the currently used indicator.
Before you return the instrument for repair work, remove the bottle
and ensure that the measuring chamber has been flushed out and
is empty.
Insert a transportation plug into the outlet pipe to prevent leakage.
Use to the enclosed error log to describe the error and return it to
us together with the instrument. If the error log is no longer
available, download it at www.heyl.de.