2.34 Active Brake System—Reduced Braking Ability
Chapter 2: Troubleshooting Procedures
2.34 Active Brake System—Reduced Braking Ability
1. Lift the shroud and squeeze the Active Brake lever. The hex rod turns.
Yes No
Replace the cable (refer to “Active Brake System” on page
4-111). If this solves the problem, go to “Final Actions” on page
2-11. Otherwise, continue to step 2.
2. Release the Active Brake lever. The hex brake lever moves to the neutral
position.
Yes No
Adjust the cable (refer to “Active Brake Adjustment” on page
4-113). If this solves the problem, go to “Final Actions” on page
2-11. Otherwise, call Technical Support.
3. Call Technical Support.
2
Hill-Rom® Transport, Procedural, and Specialty Stretchers Page 2 - 47
Service Manual (144386 REV 2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom P8000 and is the answer not in the manual?

Hill-Rom P8000 Specifications

General IconGeneral
CategoryMedical Equipment
TypeHospital Bed
ModelP8000
Weight Capacity500 lbs (227 kg)
Height AdjustmentElectric
Frame MaterialSteel
Wheel TypeLocking casters
Overall Width36 inches (91.4 cm)
Caster Diameter5 inches (12.7 cm)
Mattress Typefoam
Side Rail OptionsFull-length or half-length
CPR FunctionYes
Height Range19.5 - 30 inches (49.5 - 76.2 cm)

Related product manuals