SERVICE MANUAL
Hill-Rom®
Transport,
Procedural,
and Specialty
Stretchers
Product No. P8000, P8005, P8010, P8020, P8040, P8050
144386 REV 2

Table of Contents

Questions and Answers

  • J
    jorge77Sep 23, 2025
    Why Procedural Stretcher—Auto Contour™ Feature or BackSaver Fowler® Feature Does Not Raise or Lower Correctly in Hill-Rom P8020 Medical Equipment?
    • M
      Mark CollinsSep 23, 2025
      If the Procedural Stretcher's Auto Contour™ Feature or BackSaver Fowler® Feature does not raise or lower correctly on your Hill-Rom medical equipment, replace the cables as necessary. Alternatively, adjust the cables or replace the head-end gas cylinder as needed.
  • T
    traceyhurstAug 25, 2025
    What to do if the Hill-Rom P8020 stretcher will not lower?
    • S
      Scott ConradAug 25, 2025
      If your Hill-Rom medical equipment stretcher will not lower, you can try to adjust the release mechanism. If that doesn't work, you may need to replace the hydraulic cylinder at either the head or foot end of the stretcher.
  • A
    Annette DanielsAug 25, 2025
    Why is there reduced braking ability in my Hill-Rom Medical Equipment?
    • J
      joel00Aug 26, 2025
      To address reduced braking ability in your Hill-Rom medical equipment, ensure the brake/steer link adjustment at both ends of the stretcher is correct. If necessary, adjust the brake/steer link. If the issue persists, adjust the caster brake shoes or replace any defective casters.
  • E
    Elizabeth DixonAug 27, 2025
    What to do if the weight shown on the Hill-Rom P8020 scale display is not correct?
    • T
      Thomas RobertsAug 27, 2025
      If the weight shown on the Hill-Rom medical equipment scale display is not correct, ensure all packing materials and shims have been removed before taking a weight reading. Also, verify that the load beam cables are connected in the correct order before taking another reading.
  • A
    Anthony IrwinAug 27, 2025
    What to do if the Hill-Rom P8020 Medical Equipment stretcher will not go into neutral position?
    • M
      Michael AyersAug 27, 2025
      If your Hill-Rom medical equipment stretcher will not go into neutral position, check the brake/steer link at the head and foot ends of the stretcher for correct orientation, and adjust if necessary. Also, inspect each caster and replace any that are defective. You may also need to tighten the brake/steer link screw.
  • A
    Angela ScottSep 13, 2025
    Why Electric Stretcher—Knee Section Does Not Raise or Lower Electrically in Hill-Rom Medical Equipment?
    • H
      Heather CastroSep 14, 2025
      If the Electric Stretcher knee section on your Hill-Rom medical equipment does not raise or lower electrically, unplug the stretcher from its power source, put on an antistatic strap, and then connect the plugs firmly into position.
  • C
    Chad LeachSep 12, 2025
    Why Electric Stretcher—Back Section Does Not Raise or Lower Electrically in Hill-Rom P8020 Medical Equipment?
    • B
      Brooke KelleySep 12, 2025
      If the Electric Stretcher back section on your Hill-Rom medical equipment does not raise or lower electrically, unplug the stretcher from its power source, put on an antistatic strap, and then connect the plugs firmly into position.
  • B
    Brenda ConnerSep 12, 2025
    What causes OB/GYN Stretcher—Foot Supports Malfunction in Hill-Rom P8020?
    • P
      Pamela HayesSep 12, 2025
      If the OB/GYN Stretcher foot supports on your Hill-Rom medical equipment malfunction, replace any missing hardware. If the problem continues, replace the defective foot support gas spring or the defective foot support Mechlok®1 device.
  • P
    pgrossSep 4, 2025
    Why Procedural, Transport, Trauma, or OB/GYN Stretcher—Back Section Does Not Raise or Lower Correctly in Hill-Rom Medical Equipment?
    • K
      Kimberly BrownSep 4, 2025
      If the Procedural, Transport, Trauma, or OB/GYN Stretcher back section on your Hill-Rom medical equipment does not raise or lower correctly, replace any missing hardware. If the issue continues, replace the defective back section gas spring.
  • J
    Jay ValdezSep 2, 2025
    Why the Steering Plus™ Steering System Does Not Operate Correctly in Hill-Rom P8020 Medical Equipment?
    • C
      caroline86Sep 2, 2025
      If the Steering Plus™ Steering System on your Hill-Rom medical equipment does not operate correctly, examine the fifth wheel pivot arms and pivot bracket for correct assembly. If problems persist, remove the Steering Plus™ Steering System assembly and replace any damaged or inoperative parts, or replace the fifth wheel.