Table of Contents

Questions and Answers

  • C
    Craig JonesSep 23, 2025
    Why Procedural Stretcher—Auto Contour™ Feature or BackSaver Fowler® Feature Does Not Raise or Lower Correctly in Hill-Rom Medical Equipment?
    • L
      lucas67Sep 23, 2025
      If the Procedural Stretcher's Auto Contour™ Feature or BackSaver Fowler® Feature does not raise or lower correctly on your Hill-Rom medical equipment, replace the cables as necessary. Alternatively, adjust the cables or replace the head-end gas cylinder as needed.
  • R
    rdavisAug 25, 2025
    What to do if the Hill-Rom P8010 stretcher will not lower?
    • J
      Jeffrey WareAug 25, 2025
      If your Hill-Rom medical equipment stretcher will not lower, you can try to adjust the release mechanism. If that doesn't work, you may need to replace the hydraulic cylinder at either the head or foot end of the stretcher.
  • J
    Julia PadillaAug 25, 2025
    Why is there reduced braking ability in my Hill-Rom P8010 Medical Equipment?
    • A
      albertpetersonAug 26, 2025
      To address reduced braking ability in your Hill-Rom medical equipment, ensure the brake/steer link adjustment at both ends of the stretcher is correct. If necessary, adjust the brake/steer link. If the issue persists, adjust the caster brake shoes or replace any defective casters.
  • D
    Dawn KellyAug 27, 2025
    What to do if the weight shown on the Hill-Rom Medical Equipment scale display is not correct?
    • K
      Kathy OlsonAug 27, 2025
      If the weight shown on the Hill-Rom medical equipment scale display is not correct, ensure all packing materials and shims have been removed before taking a weight reading. Also, verify that the load beam cables are connected in the correct order before taking another reading.
  • L
    Leah GardnerAug 27, 2025
    What to do if the Hill-Rom P8010 stretcher will not go into neutral position?
    • R
      Rebecca MoyerAug 27, 2025
      If your Hill-Rom medical equipment stretcher will not go into neutral position, check the brake/steer link at the head and foot ends of the stretcher for correct orientation, and adjust if necessary. Also, inspect each caster and replace any that are defective. You may also need to tighten the brake/steer link screw.
  • R
    Robert ReedSep 13, 2025
    Why Electric Stretcher—Knee Section Does Not Raise or Lower Electrically in Hill-Rom Medical Equipment?
    • M
      Michael SchroederSep 14, 2025
      If the Electric Stretcher knee section on your Hill-Rom medical equipment does not raise or lower electrically, unplug the stretcher from its power source, put on an antistatic strap, and then connect the plugs firmly into position.
  • P
    Philip HamiltonSep 12, 2025
    Why Electric Stretcher—Back Section Does Not Raise or Lower Electrically in Hill-Rom P8010?
    • E
      erica26Sep 12, 2025
      If the Electric Stretcher back section on your Hill-Rom medical equipment does not raise or lower electrically, unplug the stretcher from its power source, put on an antistatic strap, and then connect the plugs firmly into position.
  • C
    Corey SanchezSep 12, 2025
    What causes OB/GYN Stretcher—Foot Supports Malfunction in Hill-Rom P8010?
    • C
      carolmurphySep 12, 2025
      If the OB/GYN Stretcher foot supports on your Hill-Rom medical equipment malfunction, replace any missing hardware. If the problem continues, replace the defective foot support gas spring or the defective foot support Mechlok®1 device.
  • C
    cody91Sep 4, 2025
    Why Procedural, Transport, Trauma, or OB/GYN Stretcher—Back Section Does Not Raise or Lower Correctly in Hill-Rom P8010?
    • J
      Joshua SantanaSep 4, 2025
      If the Procedural, Transport, Trauma, or OB/GYN Stretcher back section on your Hill-Rom medical equipment does not raise or lower correctly, replace any missing hardware. If the issue continues, replace the defective back section gas spring.
  • M
    Michael BrownSep 2, 2025
    Why the Steering Plus™ Steering System Does Not Operate Correctly in Hill-Rom P8010?
    • D
      Dennis ThompsonSep 2, 2025
      If the Steering Plus™ Steering System on your Hill-Rom medical equipment does not operate correctly, examine the fifth wheel pivot arms and pivot bracket for correct assembly. If problems persist, remove the Steering Plus™ Steering System assembly and replace any damaged or inoperative parts, or replace the fifth wheel.

Summary

Chapter 1: Introduction

Specifications

Details the physical, electrical, and environmental specifications of the stretchers.

Safety Tips

Offers critical safety advice for operating and servicing the stretchers to prevent injury or damage.

Chapter 2: Troubleshooting Procedures

Scale-Problem and Solution Table

A table listing common scale display problems and corresponding troubleshooting procedures (RAPs).

Function Checks

Details specific functional checks for various stretcher components to verify proper operation.

Reduced Braking Ability

Troubleshooting procedure for issues with reduced braking ability on the stretcher.

Loss of Corner Steer

Procedure to troubleshoot and resolve issues related to the loss of corner steer functionality.

Steering Plus Steering System Does Not Operate Correctly

Troubleshooting steps for when the Steering Plus™ Steering System is not functioning correctly.

Stretcher Will Not Go into Neutral Position

Guidance for diagnosing and fixing issues where the stretcher does not enter the neutral position.

Reduced Pedal Pumps

Troubleshooting steps for issues with reduced pedal pump performance on the stretcher.

Stretcher Will Not Lower

Procedure to address problems where the stretcher fails to lower correctly.

Siderail Does Not Operate Correctly

Troubleshooting steps for siderails that are not operating correctly.

Push Handle Malfunction

Steps to diagnose and resolve malfunctions related to the push handles.

Permanent IV Pole Does Not Raise or Lower Correctly

Troubleshooting for the permanent IV pole not raising or lowering correctly.

Procedural or Surgical Stretcher-Knee Section Will Not Raise or Lower

Troubleshooting steps for knee section articulation issues on specific stretcher models.

Scale Display Does Not Come On

Procedure to troubleshoot when the scale display fails to turn on.

Chapter 3: Theory of Operation

Chapter 4: Removal, Replacement, and Adjustment Procedures

Procedural (P8000), Surgical (P8010), Trauma (P8040), or OB;GYN (P8050) Stretcher-Siderail

Step-by-step instructions for removing and replacing the siderail on specified stretcher models.

Electric (P8020) Stretcher-Siderail

Detailed procedure for the removal and replacement of the siderail on the Electric (P8020) Stretcher.

Electric (P8020) Stretcher-Left or Right Siderail Cable Assembly and Upright Tube

Instructions for removing and replacing the siderail cable assembly and upright tube on the Electric (P8020) Stretcher.

Electric (P8020) Stretcher-Patient Control Switch

Procedure for removing and replacing the patient control switch on the Electric (P8020) Stretcher.

Electric (P8020) Stretcher-Nurse Control Switch

Steps for removing and replacing the nurse control switch on the Electric (P8020) Stretcher.

Electric (P8020) Stretcher-Control Board

Detailed instructions for removing and replacing the control board in the Electric (P8020) Stretcher.

Procedural (P8000), Transport (P8005), Trauma (P8040), or OB;GYN (P8050) Stretcher-Back Section Gas Spring

Procedure for removing, replacing, and adjusting the back section gas spring on specific stretcher models.

Surgical (P8010) Stretcher-Back Section Gas Spring

Instructions for removing, replacing, and adjusting the back section gas spring on the Surgical (P8010) Stretcher.

Surgical (P8010) Stretcher-Articulating Headrest Gas Spring

Procedure for removing, replacing, and adjusting the articulating headrest gas spring on the Surgical (P8010) Stretcher.

Procedural (P8000) or Surgical (P8010) Stretcher-Knee Screw Assembly

Steps for removing, replacing, and adjusting the knee screw assembly on specific stretcher models.

Electric (P8020) Stretcher-Knee Drive Assembly

Procedure for removing, replacing, and adjusting the knee drive assembly on the Electric (P8020) Stretcher.

Electric (P8020) Stretcher-Head Drive Assembly

Instructions for removing, replacing, and adjusting the head drive assembly on the Electric (P8020) Stretcher.

Electric (P8020) Stretcher-CPR Release Cable Assembly

Detailed procedure for removing, replacing, and adjusting the CPR release cable assembly.

OB;GYN Stretcher-Foot Support Mechlok Device

Procedure for removing and replacing the foot support Mechlok® device on the OB/GYN Stretcher.

OB;GYN Stretcher-Foot Support Gas Spring

Instructions for removing, replacing, and adjusting the foot support gas spring on the OB/GYN Stretcher.

Caster Brake Adjustment

Detailed steps for adjusting the caster brake shoes for proper function.

Brake;Steer Pedal Adjustment

Instructions on how to adjust the brake/steer pedal linkage for correct operation.

Hydraulic Cylinder

Procedure for removing, replacing, and adjusting the hydraulic cylinder.

Scale-Load Beam

Procedure for removing and replacing the load beams for the scale system, including calibration.

Scale-Display or Scale P.C. Board

Steps for replacing the scale display or the scale P.C. board.

Chapter 5: Parts List

Service Parts Ordering

Instructions on how to identify and order replacement parts using part numbers and model information.

Recommended Spare Parts

Provides a list of recommended spare parts for servicing five or more units.

Electric (P8020) Stretcher-Head and Knee Drive Assembly

Parts list and diagram for the head and knee drive assembly of the Electric (P8020) Stretcher.

Electric (P8020) Stretcher-Head Drive;CPR Assemblies

Parts list and diagram for the head drive and CPR assemblies of the Electric (P8020) Stretcher.

Electric (P8020) Stretcher-Control Board Assembly (120 V AC and 230 V AC)

Parts list and diagram for the control board assembly (120V AC and 230V AC) of the Electric (P8020) Stretcher.

Procedural (P8000) and Trauma (P8040) Stretchers-Siderail Assembly

Parts list and diagram for the siderail assembly on Procedural and Trauma stretchers.

Surgical (P8010) Stretcher-Siderail Assembly

Parts list and diagram for the siderail assembly on the Surgical (P8010) Stretcher.

Electric (P8020) Stretcher-Siderail Assembly

Parts list and diagram for the siderail assembly on the Electric (P8020) Stretcher.

OB;GYN (P8050) Stretcher Siderail-Assembly

Parts list and diagram for the siderail assembly on the OB/GYN (P8050) Stretcher.

Chapter 6: General Procedures

Preventive Maintenance

Importance and recommendations for establishing an effective preventive maintenance program.

Preventive Maintenance Schedule

A schedule outlining routine preventive maintenance checks for various stretcher functions.

Preventive Maintenance Checklist

A checklist designed to maintain a running history of maintenance and repair costs for stretchers.