Hill-Rom P8050 - Service Manual

Hill-Rom P8050
444 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
SERVICE MANUAL
Hill-Rom®
Transport,
Procedural,
and Specialty
Stretchers
Product No. P8000, P8005, P8010, P8020, P8040, P8050
144386 REV 2

Table of Contents

Questions and Answers

  • V
    virginiajacksonSep 23, 2025
    Why Procedural Stretcher—Auto Contour™ Feature or BackSaver Fowler® Feature Does Not Raise or Lower Correctly in Hill-Rom P8050?
    • T
      Tiffany TerrySep 23, 2025
      If the Procedural Stretcher's Auto Contour™ Feature or BackSaver Fowler® Feature does not raise or lower correctly on your Hill-Rom medical equipment, replace the cables as necessary. Alternatively, adjust the cables or replace the head-end gas cylinder as needed.
  • G
    Gary RodriguezAug 25, 2025
    What to do if the Hill-Rom Medical Equipment stretcher will not lower?
    • R
      Roy RichAug 25, 2025
      If your Hill-Rom medical equipment stretcher will not lower, you can try to adjust the release mechanism. If that doesn't work, you may need to replace the hydraulic cylinder at either the head or foot end of the stretcher.
  • C
    Cody EdwardsAug 25, 2025
    Why is there reduced braking ability in my Hill-Rom P8050?
    • K
      Kevin StokesAug 26, 2025
      To address reduced braking ability in your Hill-Rom medical equipment, ensure the brake/steer link adjustment at both ends of the stretcher is correct. If necessary, adjust the brake/steer link. If the issue persists, adjust the caster brake shoes or replace any defective casters.
  • B
    Brian ManningAug 26, 2025
    What to do if the weight shown on the Hill-Rom P8050 Medical Equipment scale display is not correct?
    • J
      Janet WardAug 27, 2025
      If the weight shown on the Hill-Rom medical equipment scale display is not correct, ensure all packing materials and shims have been removed before taking a weight reading. Also, verify that the load beam cables are connected in the correct order before taking another reading.
  • R
    roseericAug 27, 2025
    What to do if the Hill-Rom Medical Equipment stretcher will not go into neutral position?
    • P
      psharpAug 27, 2025
      If your Hill-Rom medical equipment stretcher will not go into neutral position, check the brake/steer link at the head and foot ends of the stretcher for correct orientation, and adjust if necessary. Also, inspect each caster and replace any that are defective. You may also need to tighten the brake/steer link screw.
  • J
    Jennifer LeeSep 13, 2025
    Why Electric Stretcher—Knee Section Does Not Raise or Lower Electrically in Hill-Rom P8050 Medical Equipment?
    • V
      Victoria BurnsSep 13, 2025
      If the Electric Stretcher knee section on your Hill-Rom medical equipment does not raise or lower electrically, unplug the stretcher from its power source, put on an antistatic strap, and then connect the plugs firmly into position.
  • S
    simpsoncassandraSep 12, 2025
    What causes OB/GYN Stretcher—Foot Supports Malfunction in Hill-Rom P8050 Medical Equipment?
    • T
      Timothy SweeneySep 12, 2025
      If the OB/GYN Stretcher foot supports on your Hill-Rom medical equipment malfunction, replace any missing hardware. If the problem continues, replace the defective foot support gas spring or the defective foot support Mechlok®1 device.
  • C
    Christina SmithSep 12, 2025
    Why Electric Stretcher—Back Section Does Not Raise or Lower Electrically in Hill-Rom P8050?
    • K
      Kevin HawkinsSep 12, 2025
      If the Electric Stretcher back section on your Hill-Rom medical equipment does not raise or lower electrically, unplug the stretcher from its power source, put on an antistatic strap, and then connect the plugs firmly into position.
  • P
    Patricia AndrewsSep 4, 2025
    Why Procedural, Transport, Trauma, or OB/GYN Stretcher—Back Section Does Not Raise or Lower Correctly in Hill-Rom Medical Equipment?
    • B
      Brian PalmerSep 4, 2025
      If the Procedural, Transport, Trauma, or OB/GYN Stretcher back section on your Hill-Rom medical equipment does not raise or lower correctly, replace any missing hardware. If the issue continues, replace the defective back section gas spring.
  • N
    Nathan PrestonSep 2, 2025
    Why the Steering Plus™ Steering System Does Not Operate Correctly in Hill-Rom P8050 Medical Equipment?
    • J
      Julia AdamsSep 2, 2025
      If the Steering Plus™ Steering System on your Hill-Rom medical equipment does not operate correctly, examine the fifth wheel pivot arms and pivot bracket for correct assembly. If problems persist, remove the Steering Plus™ Steering System assembly and replace any damaged or inoperative parts, or replace the fifth wheel.

Summary

Chapter 1: Introduction

Safety Tips

Provides crucial safety guidelines for operation and maintenance.

Chapter 2: Troubleshooting Procedures

Scale-Problem and Solution Table

Lists common scale display errors and their corresponding solutions.

Function Checks

Details procedures to verify the correct operation of various stretcher functions.

Specific Function Checks-Procedural Stretcher

Covers specific function checks for the Procedural Stretcher model.

Specific Function Checks-Electric Stretcher

Details function checks specific to the Electric Stretcher model.

Reduced Braking Ability

Troubleshooting steps for issues with reduced stretcher braking performance.

Stretcher Will Not Lower

Provides guidance for diagnosing and resolving issues where the stretcher will not lower.

Scale Display Does Not Come On

Troubleshooting steps for when the stretcher scale display is not functioning.

Chapter 4: Removal, Replacement, and Adjustment Procedures

Procedural (P8000), Surgical (P8010), Trauma (P8040), or OB;GYN (P8050) Stretcher-Siderail

Provides steps for removing and replacing the siderail on specific stretcher models.

Electric (P8020) Stretcher-Siderail

Details the procedure for removing and replacing the siderail on the electric stretcher.

Electric (P8020) Stretcher-Control Board

Provides instructions for removing and replacing the control board in the electric stretcher.

Scale-Load Beam

Details the procedure for removing and replacing the scale load beam.

Active Brake System

Explains the removal, replacement, and adjustment of the active brake system.

Chapter 7: Accessories