SERVICE MANUAL
Hill-Rom®
Transport,
Procedural,
and Specialty
Stretchers
Product No. P8000, P8005, P8010, P8020, P8040, P8050
144386 REV 2

Table of Contents

Questions and Answers

  • K
    Kathy DownsSep 23, 2025
    Why Procedural Stretcher—Auto Contour™ Feature or BackSaver Fowler® Feature Does Not Raise or Lower Correctly in Hill-Rom P8005?
    • J
      Joyce NorrisSep 23, 2025
      If the Procedural Stretcher's Auto Contour™ Feature or BackSaver Fowler® Feature does not raise or lower correctly on your Hill-Rom medical equipment, replace the cables as necessary. Alternatively, adjust the cables or replace the head-end gas cylinder as needed.
  • D
    Dana KimAug 25, 2025
    What to do if the Hill-Rom P8005 Medical Equipment stretcher will not lower?
    • M
      Megan ScottAug 25, 2025
      If your Hill-Rom medical equipment stretcher will not lower, you can try to adjust the release mechanism. If that doesn't work, you may need to replace the hydraulic cylinder at either the head or foot end of the stretcher.
  • J
    Jennifer HallAug 25, 2025
    Why is there reduced braking ability in my Hill-Rom P8005 Medical Equipment?
    • C
      Crystal ColemanAug 26, 2025
      To address reduced braking ability in your Hill-Rom medical equipment, ensure the brake/steer link adjustment at both ends of the stretcher is correct. If necessary, adjust the brake/steer link. If the issue persists, adjust the caster brake shoes or replace any defective casters.
  • T
    Tasha ScottAug 27, 2025
    What to do if the weight shown on the Hill-Rom Medical Equipment scale display is not correct?
    • C
      cynthiapooleAug 27, 2025
      If the weight shown on the Hill-Rom medical equipment scale display is not correct, ensure all packing materials and shims have been removed before taking a weight reading. Also, verify that the load beam cables are connected in the correct order before taking another reading.
  • J
    james65Aug 27, 2025
    What to do if the Hill-Rom P8005 Medical Equipment stretcher will not go into neutral position?
    • H
      Holly DavisAug 27, 2025
      If your Hill-Rom medical equipment stretcher will not go into neutral position, check the brake/steer link at the head and foot ends of the stretcher for correct orientation, and adjust if necessary. Also, inspect each caster and replace any that are defective. You may also need to tighten the brake/steer link screw.
  • E
    emilymitchellSep 13, 2025
    Why Electric Stretcher—Knee Section Does Not Raise or Lower Electrically in Hill-Rom P8005?
    • C
      Charles KramerSep 13, 2025
      If the Electric Stretcher knee section on your Hill-Rom medical equipment does not raise or lower electrically, unplug the stretcher from its power source, put on an antistatic strap, and then connect the plugs firmly into position.
  • J
    Jeffery DavisSep 12, 2025
    Why Electric Stretcher—Back Section Does Not Raise or Lower Electrically in Hill-Rom Medical Equipment?
    • T
      Taylor TaylorSep 12, 2025
      If the Electric Stretcher back section on your Hill-Rom medical equipment does not raise or lower electrically, unplug the stretcher from its power source, put on an antistatic strap, and then connect the plugs firmly into position.
  • A
    Angelica MathewsSep 12, 2025
    What causes OB/GYN Stretcher—Foot Supports Malfunction in Hill-Rom P8005 Medical Equipment?
    • M
      Michael AnthonySep 12, 2025
      If the OB/GYN Stretcher foot supports on your Hill-Rom medical equipment malfunction, replace any missing hardware. If the problem continues, replace the defective foot support gas spring or the defective foot support Mechlok®1 device.
  • M
    Melissa StewartSep 4, 2025
    Why Procedural, Transport, Trauma, or OB/GYN Stretcher—Back Section Does Not Raise or Lower Correctly in Hill-Rom Medical Equipment?
    • L
      Leslie JohnsonSep 4, 2025
      If the Procedural, Transport, Trauma, or OB/GYN Stretcher back section on your Hill-Rom medical equipment does not raise or lower correctly, replace any missing hardware. If the issue continues, replace the defective back section gas spring.
  • G
    gsotoSep 2, 2025
    Why the Steering Plus™ Steering System Does Not Operate Correctly in Hill-Rom P8005?
    • M
      Mrs. Kimberly MullinsSep 2, 2025
      If the Steering Plus™ Steering System on your Hill-Rom medical equipment does not operate correctly, examine the fifth wheel pivot arms and pivot bracket for correct assembly. If problems persist, remove the Steering Plus™ Steering System assembly and replace any damaged or inoperative parts, or replace the fifth wheel.

Summary

Chapter 1: Introduction

Specifications

Details the physical, electrical, environmental, and regulatory specifications of the stretchers.

Safety Tips

Provides essential safety guidelines for operating and maintaining the stretchers.

Warning and Caution Labels

Illustrates and explains various warning and caution labels found on the equipment.

Chapter 2: Troubleshooting Procedures

Getting Started

Provides an overview of how to begin troubleshooting procedures.

Initial Actions

Guides on gathering initial data from operators to identify problems.

Scale-Problem and Solution Table

A table listing common scale display errors and their solutions.

Function Checks

Procedures to verify the correct operation of various stretcher functions.

Chapter 3: Theory of Operation

Electrical System (Electric Stretcher)

Explains the electrical components and circuitry of the electric stretcher.

Scale System (Optional for the Procedural, Electric, and Trauma Stretchers)

Details the components and operation of the optional scale system.

Chapter 4: Removal, Replacement, and Adjustment Procedures

Procedural (P8000), Surgical (P8010), Trauma (P8040), or OB;GYN (P8050) Stretcher-Siderail

Procedure for removing and replacing the siderail on specific stretcher models.

Electric (P8020) Stretcher-Siderail

Procedure for removing and replacing the siderail on the electric stretcher.

Transport (P8005) Stretcher-Siderail Latch

Procedure for removing and replacing the siderail latch.

Transport (P8005) Stretcher-Siderail Tube

Procedure for removing and replacing the siderail tube.

Procedural (P8000), Surgical (P8010), Electric (P8020), Trauma (P8040), or OB;GYN (P8050) Stretcher-Siderail Latch

Procedure for removing/replacing siderail latch on multiple stretcher models.

Procedural (P8000), Surgical (P8010), Electric (P8020), Trauma (P8040), or OB;GYN (P8050) Stretcher-Siderail Tubes

Procedure for removing/replacing siderail tubes on multiple stretcher models.

OB;GYN (P8050) Stretcher-Patient Grip Handle

Procedure for removing and replacing the patient grip handle.

Electric (P8020) Stretcher-Patient Control Switch

Procedure for removing and replacing the patient control switch.

Electric (P8020) Stretcher-Nurse Control Switch

Procedure for removing and replacing the nurse control switch.

Electric (P8020) Stretcher-Control Board

Procedure for removing and replacing the control board.

Electric Stretcher-Potentiometer

Procedure for removing and replacing the potentiometer.

Electric Stretcher-Power Cord Assembly

Procedure for removing and replacing the power cord assembly.

Surgical (P8010) Stretcher-Back Section Gas Spring

Procedure for removing and replacing the back section gas spring.

Surgical (P8010) Stretcher-Articulating Headrest Gas Spring

Procedure for removing and replacing the articulating headrest gas spring.

Procedural (P8000) or Surgical (P8010) Stretcher-Knee Screw Assembly

Procedure for removing and replacing the knee screw assembly.

Electric (P8020) Stretcher-Knee Drive Assembly

Procedure for removing and replacing the knee drive assembly.

Electric (P8020) Stretcher-Head Drive Assembly

Procedure for removing and replacing the head drive assembly.

Electric (P8020) Stretcher-CPR Release Cable Assembly

Procedure for removing and replacing the CPR release cable assembly.

Upper Frame and Base Shroud

Procedure for removing and replacing the upper frame and base shroud.

OB;GYN (P8050) Stretcher-Upper Frame Catch Shroud

Procedure for removing and replacing the upper frame catch shroud.

OB;GYN Stretcher-Foot Support Mechlok Device

Procedure for removing and replacing the foot support Mechlok® device.

OB;GYN Stretcher-Foot Support Gas Spring

Procedure for removing and replacing the foot support gas spring.

Caster

Procedure for removing and replacing a stretcher caster.

Chapter 5 Parts List

Service Parts Ordering

Information on how to identify and order replacement parts.

Chapter 6 General Procedures

Cleaning and Care

Provides instructions for cleaning and caring for the stretcher.

Preventive Maintenance

Emphasizes the importance of a regular maintenance program.

Preventive Maintenance Schedule

Outlines the scheduled preventive maintenance tasks for the stretchers.

Chapter 7 Accessories

Stretcher Accessories

Lists all available stretcher accessories with product numbers.

Hill-Rom P8005 Specifications

General IconGeneral
Weight Capacity500 lbs (227 kg)
ModelP8005
ManufacturerHill-Rom
Trendelenburg/Reverse TrendelenburgYes
Mattress CompatibilityCompatible with Hill-Rom mattresses
Power Requirements120V AC, 60Hz
Side Rail OptionsFull, Half, or No Side Rails