3550 Battery HiTESTER Series
5
Q7. I want to externally control
EXT.HOLD and EXT.MEMO.
Where is the circuit common
point? Is it possible to perform
control with 0/5 V (CMOS
level)?
A7. Refer to the circuit configuration of EXT.HOLD
and EXT.MEMO shown below.The internal
circuit operates at 5 V. Control at the CMOS
level is therefore possible.
Q8. Does the printout differ,
depending on the software
version of the 9203?
A8. The software version and compatible models
are listed below.
Ver.1.00:3227
Ver.1.01:3550, 3227
Ver.1.10:3550, 3551, 3227
When version 1.10 is used and the 3550 is
connected, "NG" is displayed as "- - ".
When the 3551 is connected, printout is as
explained in the manual. When version 1.01
is used and the 3551 is connected, the 3 m
range cannot be printed. For details,
please contact the HIOKI Head Office.
3 Comparator…………...………………………………………………
Q1. Is it possible for HIOKI to
research the battery data and
write the comparator values to
memory?
A1. Although HIOKI is engaged in battery
charging and discharging tests, the data
generated by these tests is intended only for
reference purposes. Users must provide the
data on individual batteries.
Q2. Is it possible to increase the
number of comparator?
A2. Yes. It is possible to increase the number of
comparator numbers up to 30(3550 and 3555
but 3551 is not possible). Such cases are
handled as special orders.
Q3. Is it possible to input upper and
lower limit values for a voltage
comparator?
A3. This is not possible with the 3550 series.
Inspections conducted by the battery
manufacturer when the batteries are shipped
from the factor are performed with upper and
lower limit values. However, if the voltage
is higher than normal during maintenance