76
Setting Threshold Values for the Comparator
Comparison table for the comparator
The instrument will show a result and produce beeps as shown in
the following table:
Resistance
(low)
Resistance
(medium)
Resistance
(high)
Voltage (high) PASS WARNING FAIL
Voltage warning threshold
Voltage (low) WARNING WARNING FAIL
Resistance
warning threshold
Resistance
rejection value
The boundary conditions are as follows:
Resistance
PASS
≤
Resistance
warning
threshold
<
Resistance
WARNING
≤
Resistance
rejection value
<
Resistance
FAIL
Voltage
WARNING
<
Voltage
warning
threshold
≤
Voltage
PASS
HIOKI BT3554F961-00HIOKI BT3554F961-00