EasyManua.ls Logo

Hioki MO HiTESTER 3452 - Page 76

Hioki MO HiTESTER 3452
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
―――――――――――――――――――――――――――
Chapter 3 Specifications
――――――――――――――――――――――――
A
llowance of resistance measurement
1st effective measuremen
t
range (MΩ)
0
.01 - 5
0
.01 - 5
0
.02 - 10
±5% of scale reading
2nd effective
measurement range
(MΩ)
5 - 10,
0
.005 -
0
.01
5 - 10,
0
.005 -
0
.01
10 - 20,
0
.01 -
0
.02
±10% of scale reading
0MΩ, scale 0.7% of 100 V scale length
(
When needle centerline is
w
ithin width of scale markings
a
t the extreme ends of the scale.)
M
easurement terminal voltage characteristic
Open-circuit voltage
(when no load is applied)
1 to 1.2 times of rated
measurement voltage (open
circuit terminal voltage)
Lower limit measurement
resistance value to be
maintained rated
measurement voltage (MΩ)
0
.025
0
.05
0
.1
Rated measurement current
(current to be maintained
rated measurement voltage)
1 - 1.2 mA
Short circuit current 1.2 mA max
A
C voltage measurement
AC voltage scale
(50/60 Hz) and allowance
0 to 150 V, ±5% (=±7.5 V)
of maximum scale value
Input resistance 30 kΩ min
M
aximum rated voltage
t
o earth
150 V AC
I
nput error protection for
1
0 seconds
200 V AC
(overvoltage protection)

Related product manuals