Stack Gas Analysis System
ENDA-5000
Instruction Manual
CODE:GZ9100497217D

Table of Contents

Questions and Answers

  • M
    Matthew BowersAug 15, 2025
    What to do if calibration is not possible on Horiba Measuring Instruments?
    • D
      David MartinezAug 15, 2025
      If calibration is not possible on your Horiba Measuring Instruments, there are several potential causes and solutions: * If the calibration cylinder is empty, replace it with a new one. * If the wrong calibration gas type is selected, set the correct gas type. * If the calibration concentration set value and the cylinder concentration value are different, set the calibration concentration again and perform the zero/span calibration again. * If the calibration gas flow is too low, adjust the flow to the specified rate. * If the device is in a key-lock state, input the password to release it.
  • R
    Rebecca FosterAug 18, 2025
    What to do if I have a Zero Calibration Alarm on my horiba ENDA-5000 Measuring Instruments?
    • N
      Nicholas MillerAug 19, 2025
      If you encounter a Zero Calibration Alarm on your Horiba Measuring Instruments: * Ensure the correct gas type is selected; if not, redo the calibration with the appropriate gas. * If the calibration gas residual pressure is within 1 MPa, replace the cylinder and recalibrate. * Verify the flow is within the specified level, adjust if necessary, and recalibrate. * If the issue persists, it is recommended to contact HORIBA.
  • P
    Phillip JohnstonAug 21, 2025
    What to do if I have a Span Calibration Alarm on my Horiba Measuring Instruments?
    • F
      fbrownAug 21, 2025
      If you encounter a Span Calibration Alarm on your Horiba Measuring Instruments: * Ensure the correct gas type is selected; if not, redo the calibration with the appropriate gas. * If the calibration gas residual pressure is within 1 MPa, replace the cylinder and recalibrate. * Verify the flow is within the specified level, adjust if necessary, and recalibrate. * If the issue persists, it is recommended to contact HORIBA.
  • K
    Kirsten RobinsonAug 25, 2025
    What to do if I have a Zero Calibration Caution on my horiba ENDA-5000?
    • J
      John BruceAug 25, 2025
      If you encounter a Zero Calibration Caution on your Horiba Measuring Instruments: * Ensure the correct gas type is selected; if not, redo the calibration with the appropriate gas. * If the calibration gas residual pressure is within 1 MPa, replace the cylinder and recalibrate. * Verify the flow is within the specified level, adjust if necessary, and recalibrate. * If the issue persists, it is recommended to contact HORIBA.
  • P
    Pamela TylerAug 27, 2025
    What to do if I have a Span Calibration Caution on my horiba ENDA-5000 Measuring Instruments?
    • B
      Bailey SheltonAug 27, 2025
      If you encounter a Span Calibration Caution on your Horiba Measuring Instruments: * Ensure the correct gas type is selected; if not, redo the calibration with the appropriate gas. * If the calibration gas residual pressure is within 1 MPa, replace the cylinder and recalibrate. * Verify the flow is within the specified level, adjust if necessary, and recalibrate. * If the issue persists, it is recommended to contact HORIBA.
  • S
    Steven VargasAug 30, 2025
    What causes the Electric Cooler Temperature Alarm in horiba ENDA-5000?
    • S
      Sean AguirreAug 30, 2025
      The Electric Cooler Temperature Alarm in Horiba Measuring Instruments can be caused by cooling unit deterioration or power supply breakage. If the power lamp is off or the cooling fan stops, contact HORIBA. If the cooling fan or radiation fin is dirty, clean it. If the alarm persists, contact HORIBA.
  • R
    Robert WashingtonSep 2, 2025
    What should I do if the specified flow is not flowed in horiba ENDA-5000?
    • C
      Catherine HicksSep 2, 2025
      If the specified flow (0.6 L/min ± 0.1 L/min) is not flowing in your Horiba Measuring Instruments, adjust the flow to the specified rate using the flow adjusting needle valve.
  • K
    Kelly MuellerSep 5, 2025
    What causes Temperature in Cabinet Alarm (optional) in Horiba Measuring Instruments?
    • K
      Kathleen DavisSep 6, 2025
      The Temperature in Cabinet Alarm (optional) in Horiba Measuring Instruments can be triggered by temperature fluctuations outside the cabinet. If the temperature is high, consider installing a shield. If the temperature is low, contact HORIBA.
  • R
    Rodney CurrySep 8, 2025
    What to do if gas residual pressure decreased (optional) in horiba ENDA-5000?
    • V
      Valerie HinesSep 8, 2025
      If the gas residual pressure in your Horiba Measuring Instruments decreases (optional), and the residual pressure is within 1 MPa, replace the cylinder with a new one.
  • F
    Francis ShawSep 12, 2025
    What to do if the sample outlet is clogged in Horiba Measuring Instruments?
    • W
      wilsonjonathanSep 12, 2025
      If the sample outlet of your Horiba Measuring Instruments is clogged, release it into the air or prepare a tubing layout that is free from backpressure.