Questions and Answers

  • J
    jacksonstephanieAug 13, 2025
    What to do if I get a Turbidity sensor wiper motor error on my horiba U-51 Measuring Instruments?
    • W
      Wanda TownsendAug 13, 2025
      If you encounter a turbidity sensor wiper motor error, it indicates that the turbidity sensor wiper is not operating. First, press the ESC key and check for any obstacles near the wiper. Then, attempt the measurement again. If the error continues, the motor likely needs replacement, and you should contact your nearest sales outlet to have the sensor probe repaired.
  • J
    James CoxAug 16, 2025
    What to do if I get a Probe EEPROM error on my horiba U-51?
    • D
      David CisnerosAug 16, 2025
      If you receive a Probe EEPROM error, it is due to an internal IC failure. First, press the ESC key and redo the operation. If the error persists, turn the power OFF, then restart the system (note that the current data will not be saved). If the error continues, contact your nearest sales outlet to have the display repaired.
  • D
    Daniel PayneAug 20, 2025
    What to do if I get a Turbidity sensor light source error on my Horiba Measuring Instruments?
    • T
      tlewisAug 20, 2025
      If you encounter a turbidity sensor light source error, it means there is a failure with the turbidity sensor light source. Turn the power OFF, then wipe off any water droplets on the probe. Remove the turbidity sensor and check for water droplets around the turbidity sensor connector, then mount the sensor again. If the error persists, replace the turbidity sensor.
  • K
    Kyle KingAug 25, 2025
    What causes a Probe capacitor error on horiba U-51 Measuring Instruments?
    • K
      Kevin OlsonAug 25, 2025
      A Probe capacitor error is caused by low battery voltage or internal IC failure. To address this, turn the power OFF and replace the display's batteries. If the error persists, contact your nearest sales outlet to have the sensor probe repaired.
  • R
    Richard Andrews MDAug 28, 2025
    How to fix Probe EEPROM error/User on horiba U-51 Measuring Instruments?
    • R
      Rebecca AndersonAug 28, 2025
      A Probe EEPROM error/User is caused by an internal IC failure. Turn the power OFF, then restart the system. If the error persists, initialize the system from the 'System' menu. If the error still persists, contact your nearest sales outlet to have the sensor probe repaired.
  • J
    jgouldAug 31, 2025
    How to fix Probe EEPROM error/Factory on Horiba Measuring Instruments?
    • S
      ssmithAug 31, 2025
      A Probe EEPROM error/Factory is caused by an internal IC failure. Turn the power OFF, then restart the system. If the error persists, initialize the system from the 'System' menu. If the error still persists, contact your nearest sales outlet to have the sensor probe repaired.
  • M
    Michelle NealSep 3, 2025
    What to do if I get a Probe board error on my horiba U-51 Measuring Instruments?
    • J
      Jonathan BrownSep 4, 2025
      If you encounter a probe board error, it indicates a probe board failure. Turn the power OFF and contact your nearest sales outlet to have the sensor probe repaired.
  • C
    coreybrownSep 8, 2025
    What causes a Probe ADC error on horiba U-51?
    • L
      lucasmatthewSep 8, 2025
      A Probe ADC error is caused by an internal IC failure. Contact your nearest sales outlet to have the sensor probe repaired.

Summary

Safety Information

Hazard Classification and Symbols

Explains hazard classification levels (DANGER, WARNING, CAUTION) and warning symbols.

Safety Precautions

General Safety Warnings

Provides critical warnings about product modification, glass sensors, and chemical solutions.

Product Handling Information

Control Unit Information

Device Information

Measurement Parameters

Lists all measurable parameters and the U-50 models that support them.

Settings and Configuration

Sensor Selection

Details how to select and display desired measurement parameters on the control unit.

Compensation Functions

Explains compensation functions for DO, conductivity, and TDS for accuracy.

Basic Operation

System Setup

Explains how to install and replace the batteries in the control unit.

Attaching Sensors

Provides instructions on how to attach the pH, ORP, and DO sensors to the probe.

Sensor Conditioning

Explains the conditioning procedure for sensors before first use or after long storage.

Calibration Procedures

Auto Calibration

Details the process for performing automatic calibration of multiple sensors.

Manual Calibration

Describes how to calibrate each sensor individually.

Maintenance

Troubleshooting

Error Displays

Lists common error messages, their causes, and recommended solutions.

Specifications

horiba U-51 Specifications

General IconGeneral
Brandhoriba
ModelU-51
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals