Questions and Answers

  • B
    bgonzalezAug 12, 2025
    What to do if I get a Turbidity sensor wiper motor error on my Horiba Measuring Instruments?
    • S
      Sherri TurnerAug 12, 2025
      If you encounter a turbidity sensor wiper motor error, it indicates that the turbidity sensor wiper is not operating. First, press the ESC key and check for any obstacles near the wiper. Then, attempt the measurement again. If the error continues, the motor likely needs replacement, and you should contact your nearest sales outlet to have the sensor probe repaired.
  • M
    Michael YoderAug 16, 2025
    What to do if I get a Probe EEPROM error on my horiba U-52?
    • T
      Todd CooperAug 16, 2025
      If you receive a Probe EEPROM error, it is due to an internal IC failure. First, press the ESC key and redo the operation. If the error persists, turn the power OFF, then restart the system (note that the current data will not be saved). If the error continues, contact your nearest sales outlet to have the display repaired.
  • S
    Spencer WilsonAug 19, 2025
    What to do if I get a Turbidity sensor light source error on my Horiba Measuring Instruments?
    • A
      Ashley MayAug 19, 2025
      If you encounter a turbidity sensor light source error, it means there is a failure with the turbidity sensor light source. Turn the power OFF, then wipe off any water droplets on the probe. Remove the turbidity sensor and check for water droplets around the turbidity sensor connector, then mount the sensor again. If the error persists, replace the turbidity sensor.
  • J
    Jacqueline AndersonAug 23, 2025
    What causes a Probe capacitor error on horiba U-52?
    • D
      deanna98Aug 23, 2025
      A Probe capacitor error is caused by low battery voltage or internal IC failure. To address this, turn the power OFF and replace the display's batteries. If the error persists, contact your nearest sales outlet to have the sensor probe repaired.
  • B
    Brandon ChandlerAug 27, 2025
    How to fix Probe EEPROM error/User on horiba U-52?
    • M
      Mckenzie WatsonAug 27, 2025
      A Probe EEPROM error/User is caused by an internal IC failure. Turn the power OFF, then restart the system. If the error persists, initialize the system from the 'System' menu. If the error still persists, contact your nearest sales outlet to have the sensor probe repaired.
  • E
    Eric DavisAug 30, 2025
    How to fix Probe EEPROM error/Factory on Horiba Measuring Instruments?
    • C
      cody33Aug 31, 2025
      A Probe EEPROM error/Factory is caused by an internal IC failure. Turn the power OFF, then restart the system. If the error persists, initialize the system from the 'System' menu. If the error still persists, contact your nearest sales outlet to have the sensor probe repaired.
  • J
    Jennifer FrazierSep 3, 2025
    What to do if I get a Probe board error on my horiba U-52 Measuring Instruments?
    • D
      dgonzalesSep 3, 2025
      If you encounter a probe board error, it indicates a probe board failure. Turn the power OFF and contact your nearest sales outlet to have the sensor probe repaired.
  • A
    Alexander LeeSep 7, 2025
    What causes a Probe ADC error on horiba U-52?
    • V
      Vanessa NguyenSep 7, 2025
      A Probe ADC error is caused by an internal IC failure. Contact your nearest sales outlet to have the sensor probe repaired.