    Courtney SmithAug 16, 2025
    What to do if the display is blinking on my horiba HR-200RT?
    • S
      Scott MoyerAug 16, 2025
      If your Horiba Measuring Instruments display is blinking and the value is outside the measurement range, it could be due to a high concentration. Verify that the concentration is within the measurement range by performing manual analysis. If the concentration is within the measurement range, contact HORIBA Advanced Techno. Another cause could be a temperature sensor error. In this case, perform a temperature calibration.
  • N
    Nathan SmithAug 22, 2025
    Why is there fluctuation in the measured value of my horiba HR-200RT Measuring Instruments?
    • E
      Elizabeth OrrAug 22, 2025
      If you're experiencing fluctuation in the measured value of your Horiba Measuring Instruments, there are several potential causes. Air bubbles may be forming in the sample, so prevent their formation. A sudden change in the liquid temperature can also cause this issue, so install the instrument in a location with stable temperature. Ensure that the pH fluctuation is moderate. The electrical conductivity should be 8 mS/m or above. Adjust the flow rate to be within the specification range. Set a higher value for dF if the moving average count is low. Supply the specified amount of beads if an excessive amount is supplied. Also, check and tighten the terminal block screw securely and remove any humidity or contamination from the terminal block.
  • N
    Norma AndersonAug 25, 2025
    How to troubleshoot a horiba HR-200RT Measuring Instruments that does not turn ON?
    • J
      Jeremy HarveyAug 26, 2025
      If your Horiba Measuring Instruments does not turn on, the power supply voltage may be outside the rated voltage range, so check the power supply voltage. Also, check the power distribution line for any wiring errors. If these steps don't resolve the issue, there may be an internal error of the converter, in which case you should contact HORIBA Advanced Techno.
  • R
    Robert LeachAug 28, 2025
    What does E-61 error mean on my Horiba Measuring Instruments?
    • H
      Heather KirkAug 28, 2025
      If your Horiba Measuring Instruments displays 'E-61 (Flow rate error)', it indicates that the sample flow rate is not appropriate. Check the flow rate and adjust it accordingly.
  • J
    Jack MorenoAug 31, 2025
    What does E-25 error mean on my horiba HR-200RT Measuring Instruments?
    • L
      Louis SmithAug 31, 2025
      If your Horiba Measuring Instruments displays 'E-25 (Temperature is outside the measurement range)', it means the temperature is below ?5°C or above 55°C. Adjust the liquid temperature to be within the range of ?5°C to 55°C.
  • A
    andrewjohnsonSep 3, 2025
    What does E-62 error mean on my horiba HR-200RT Measuring Instruments?
    • D
      debramooreSep 3, 2025
      If your Horiba Measuring Instruments displays 'E-62 (pH correction error)', it means that the pH value of the measured liquid deviates from the range of 5.8 pH to 8.6 pH. Adjust the pH value of the sample within the range from 5.8 pH to 8.6 pH.
  • A
    Anthony CooperSep 6, 2025
    What does E-22 error mean on my horiba HR-200RT Measuring Instruments?
    • L
      Lori BeckerSep 6, 2025
      If your Horiba Measuring Instruments displays 'E-22 (Short circuit of temperature sensor)', it indicates that the lead wire of the temperature sensor is short-circuited. Check that the wiring between T and T is not short-circuited.
  • O
    Omar StricklandSep 9, 2025
    What does E-21 error mean on my Horiba Measuring Instruments?
    • C
      Cody JoyceSep 9, 2025
      If your Horiba Measuring Instruments displays 'E-21 (Disconnection of temperature sensor)', it indicates that the lead wire of the temperature sensor is disconnected. Check that the wiring between T and T is not open.
  • Z
    zachary77Sep 10, 2025
    What to do if my horiba HR-200RT indicates an abnormal value?
    • C
      cwatsonSep 12, 2025
      If your Horiba Measuring Instruments shows an abnormal value, several factors could be at play. Start by checking for electrode contamination and clean the electrode, polishing it with the provided abrasive paper. Next, inspect the connector cable for any disconnections or short circuits. Also, check the terminal block for contact failure. Ensure the sample flow rate is within the specified range, and verify that the wiring is correctly connected to the terminal block.
  • J
    Jeremy AlexanderSep 12, 2025
    What does E-75 error mean on my horiba HR-200RT?
    • T
      timothy37Sep 12, 2025
      If your Horiba Measuring Instruments shows 'E-75 Sensor error', it indicates that the sensor is damaged. Check the wire connection and correct it to a normal condition.

Summary

Preface

Warranty and responsibility

Details HORIBA's warranty terms, exclusions, and limitations of liability for the product.

Safety Information

Hazard classification and warning symbols

Explains hazard classification (DANGER, WARNING, CAUTION) and warning symbols.

Safety precautions

Lists critical safety measures for operation, including electrical and chemical handling.

Installation

Installation environment

Specifies the required environmental conditions for installing the converter and detector unit.

Piping

Provides instructions and precautions for sample inlet and outlet piping.

Wire connection

Details terminal block connections for power supply and signal wiring.

Operation and Measurement

Preparation for operation

Lists essential steps before operating the instrument for the first time.

Meas mode

Details how to operate in the primary measurement and control mode.

Setup menu

Allows setting measurement conditions, display, calibration, and transmission outputs.

Cal menu

Enables calibration of residual chlorine and temperature readings.

Maintenance and Troubleshooting

Periodic inspection list

Lists inspection items for the residual chlorine detector unit and recommended cycles.

Problems with measurement

Lists common measurement issues, their causes, and remedial actions.

Error codes

Lists various error codes, their descriptions, and occurrence.

Data and Disposal

Specifications

Lists technical specifications for the converter unit.

Parts list

Lists available options and replacement parts for the instrument.

Disposing of the instrument

Provides instructions for proper product disposal according to local laws and regulations.

horiba HR-200RT Specifications

General IconGeneral
Brandhoriba
ModelHR-200RT
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals