  • S
    sotomichaelAug 12, 2025
    What to do if I get a Turbidity sensor wiper motor error on my Horiba Measuring Instruments?
    • D
      Daniel MartinAug 12, 2025
      If you encounter a turbidity sensor wiper motor error, it indicates that the turbidity sensor wiper is not operating. First, press the ESC key and check for any obstacles near the wiper. Then, attempt the measurement again. If the error continues, the motor likely needs replacement, and you should contact your nearest sales outlet to have the sensor probe repaired.
  • J
    Jordan AdamsAug 16, 2025
    What to do if I get a Probe EEPROM error on my Horiba Measuring Instruments?
    • D
      Dave CookAug 16, 2025
      If you receive a Probe EEPROM error, it is due to an internal IC failure. First, press the ESC key and redo the operation. If the error persists, turn the power OFF, then restart the system (note that the current data will not be saved). If the error continues, contact your nearest sales outlet to have the display repaired.
  • D
    Dr. Patrick Fitzgerald IVAug 20, 2025
    What to do if I get a Turbidity sensor light source error on my horiba U-52G Measuring Instruments?
    • K
      Kimberly WalkerAug 20, 2025
      If you encounter a turbidity sensor light source error, it means there is a failure with the turbidity sensor light source. Turn the power OFF, then wipe off any water droplets on the probe. Remove the turbidity sensor and check for water droplets around the turbidity sensor connector, then mount the sensor again. If the error persists, replace the turbidity sensor.
  • L
    Lisa AndersonAug 25, 2025
    What causes a Probe capacitor error on horiba U-52G?
    • M
      Mr. Richard SawyerAug 25, 2025
      A Probe capacitor error is caused by low battery voltage or internal IC failure. To address this, turn the power OFF and replace the display's batteries. If the error persists, contact your nearest sales outlet to have the sensor probe repaired.
  • P
    penningtonkatieAug 27, 2025
    How to fix Probe EEPROM error/User on Horiba Measuring Instruments?
    • S
      simpsondeannaAug 27, 2025
      A Probe EEPROM error/User is caused by an internal IC failure. Turn the power OFF, then restart the system. If the error persists, initialize the system from the 'System' menu. If the error still persists, contact your nearest sales outlet to have the sensor probe repaired.
  • C
    campbellkyleAug 31, 2025
    How to fix Probe EEPROM error/Factory on horiba U-52G Measuring Instruments?
    • V
      Victor WoodardAug 31, 2025
      A Probe EEPROM error/Factory is caused by an internal IC failure. Turn the power OFF, then restart the system. If the error persists, initialize the system from the 'System' menu. If the error still persists, contact your nearest sales outlet to have the sensor probe repaired.
  • E
    Eric MorganSep 3, 2025
    What to do if I get a Probe board error on my Horiba Measuring Instruments?
    • H
      Hunter WashingtonSep 4, 2025
      If you encounter a probe board error, it indicates a probe board failure. Turn the power OFF and contact your nearest sales outlet to have the sensor probe repaired.
  • D
    Daniel CookSep 7, 2025
    What causes a Probe ADC error on Horiba Measuring Instruments?
    • M
      mahoneysarahSep 7, 2025
      A Probe ADC error is caused by an internal IC failure. Contact your nearest sales outlet to have the sensor probe repaired.

Summary

Preface

Warranty and Responsibility

Manufacturer's warranty terms, limitations, and responsibilities for product defects.

For Your Safety

Hazard Classification and Symbols

Explains hazard levels (DANGER, WARNING, CAUTION) and associated symbols.

Safety Precautions

General Safety Warnings

Important safety instructions for product operation and handling.

Product Handling Information

Operational Precautions

Guidelines for safe and proper product usage and maintenance.

Device Information

Basic Operation

System Setup

Instructions for initial system setup, including battery installation.

Attaching Sensors

Settings

Compensation

Calibration

Auto Calibration Procedure

Steps for performing automatic multi-parameter calibration using standard solutions.

Manual Calibration

Conductivity Calibration

Maintenance

Troubleshooting

Error Displays and Causes

Lists common errors, their causes, and troubleshooting solutions.

horiba U-52G Specifications

General IconGeneral
ParameterspH, ORP, Conductivity, Salinity, TDS, Temperature, Turbidity, Dissolved Oxygen
pH Range0.00 to 14.00 pH
pH Resolution0.01 pH
ORP Resolution1 mV
TDS Range0 to 100 g/L
Temperature Resolution0.1 °C
WaterproofIP67
Measurement PrincipleElectrochemical, Optical
Salinity Range0.0 to 70.0 ppt
Salinity Resolution0.1 PSU
Turbidity Resolution0.1 NTU