  • S
    Shawn BrownAug 13, 2025
    What to do if I get a Turbidity sensor wiper motor error on my horiba U-53?
    • B
      brandon49Aug 13, 2025
      If you encounter a turbidity sensor wiper motor error, it indicates that the turbidity sensor wiper is not operating. First, press the ESC key and check for any obstacles near the wiper. Then, attempt the measurement again. If the error continues, the motor likely needs replacement, and you should contact your nearest sales outlet to have the sensor probe repaired.
  • W
    Wesley McintoshAug 17, 2025
    What to do if I get a Probe EEPROM error on my horiba U-53?
    • D
      Danielle ScottAug 17, 2025
      If you receive a Probe EEPROM error, it is due to an internal IC failure. First, press the ESC key and redo the operation. If the error persists, turn the power OFF, then restart the system (note that the current data will not be saved). If the error continues, contact your nearest sales outlet to have the display repaired.
  • R
    Rita DunnAug 20, 2025
    What to do if I get a Turbidity sensor light source error on my horiba U-53?
    • M
      marissa35Aug 21, 2025
      If you encounter a turbidity sensor light source error, it means there is a failure with the turbidity sensor light source. Turn the power OFF, then wipe off any water droplets on the probe. Remove the turbidity sensor and check for water droplets around the turbidity sensor connector, then mount the sensor again. If the error persists, replace the turbidity sensor.
  • M
    Megan CummingsAug 25, 2025
    What causes a Probe capacitor error on Horiba Measuring Instruments?
    • A
      Andrea HicksAug 25, 2025
      A Probe capacitor error is caused by low battery voltage or internal IC failure. To address this, turn the power OFF and replace the display's batteries. If the error persists, contact your nearest sales outlet to have the sensor probe repaired.
  • A
    Alexandra NelsonAug 28, 2025
    How to fix Probe EEPROM error/User on Horiba Measuring Instruments?
    • M
      Mr. John Rodriguez IVAug 28, 2025
      A Probe EEPROM error/User is caused by an internal IC failure. Turn the power OFF, then restart the system. If the error persists, initialize the system from the 'System' menu. If the error still persists, contact your nearest sales outlet to have the sensor probe repaired.
  • D
    dawn93Aug 31, 2025
    How to fix Probe EEPROM error/Factory on Horiba Measuring Instruments?
    • P
      perrydonaldSep 1, 2025
      A Probe EEPROM error/Factory is caused by an internal IC failure. Turn the power OFF, then restart the system. If the error persists, initialize the system from the 'System' menu. If the error still persists, contact your nearest sales outlet to have the sensor probe repaired.
  • D
    dustinstricklandSep 4, 2025
    What to do if I get a Probe board error on my horiba U-53 Measuring Instruments?
    • E
      Elizabeth JohnsonSep 4, 2025
      If you encounter a probe board error, it indicates a probe board failure. Turn the power OFF and contact your nearest sales outlet to have the sensor probe repaired.
  • M
    Michael HinesSep 8, 2025
    What causes a Probe ADC error on horiba U-53 Measuring Instruments?
    • L
      Lisa CantrellSep 8, 2025
      A Probe ADC error is caused by an internal IC failure. Contact your nearest sales outlet to have the sensor probe repaired.

Summary

For Your Safety

Hazard Classification and Warning Symbols

Explains hazard classification levels (DANGER, WARNING, CAUTION) and warning symbols.

Safety Label List

Lists product labels and their corresponding hazard types and ID numbers.

Falling Hazard Warning

Details a warning about falling and how to avoid it, with instructions.

Safety Precautions

General Safety Warnings and Handling Precautions

Warnings about disassembly, glass sensor handling, ingestion, and USB communication safety.

Product Handling Information

About this Unit

Device Information

Basic Operation

Settings

Compensation Overview and Salinity (DO) Configuration

Explains compensation functions and how to configure salinity compensation for DO.

Atmospheric Pressure and Temperature Compensation

How to compensate DO for atmospheric pressure and conductivity for temperature.

Calibration

Calibration Notes and Auto Calibration Procedure

Important notes before calibration and steps for the auto calibration process.

Manual Calibration Procedures (Temp, pH)

Steps for manual calibration of temperature and pH sensors.

Manual Calibration Procedures (ORP, COND)

Steps for manual calibration of ORP and conductivity sensors.

Manual Calibration Procedures (TURB)

Guide for turbidity calibration, including standard solution preparation.

Manual Calibration Procedures (DO)

Guide for dissolved oxygen calibration, including solution preparation.

Measurement

Data Operations

Sensor Information

Maintenance

Troubleshooting

Error Displays and Solutions

Lists common errors, their causes, and solutions for the unit.

Zero-Point Calibration Errors

Lists errors specifically related to zero-point calibration failures.

Span Calibration, Stability, Turbidity, and Wet Check Errors

Lists errors for span, stability, turbidity, and wet check issues.

Specifications

Measurement Specifications (pH, DO, Conductivity, Salinity, TDS, SG)

Technical specifications for pH, DO, Conductivity, Salinity, TDS, and SG measurements.

Measurement Specifications (Temperature, Turbidity, Depth, ORP)

Technical specifications for Temperature, Turbidity, Depth, and ORP measurements.

Reference

DO Measurement: Principle, Conditioning, Salinity, and Atmospheric Pressure Compensation

Explains DO principle, conditioning, salinity, and pressure compensation.

Turbidity (TURB) Measurement: Principle and Standard Solution

Explains measurement methods and standard solutions for turbidity.

horiba U-53 Specifications

General IconGeneral
Brandhoriba
ModelU-53
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals